Conference Paper: Cavitation in high suction tensiometers: Effect of temperature, time of use, and stone drying
Title | Cavitation in high suction tensiometers: Effect of temperature, time of use, and stone drying |
---|---|
Authors | |
Issue Date | 2012 |
Citation | 5th Asia-Pacific Conference on Unsaturated Soils 2012, 2012, v. 1, p. 385-391 How to Cite? |
Abstract | In high suction tensiometers, cavitation occurs when the tension in the water inside the device exceeds a certain threshold. The threshold corresponds to the formation of air bubbles either by air entry into or formation within the porous stone or reservoir. The standard procedure to inhibit air bubble formation in the device is to flood under vacuum and apply water at high pressure to force any remaining air bubbles to dissolve. A series of factors are, however, likely to influence the cavitation behaviour of high suction tensiometers. This paper demonstrates that: (1) there are no precursory phenomena leading to cavitation, (2) higher pressures are not essential for saturation, (3) cooler temperatures and tensiometer usage decrease the pore water pressure at cavitation, and (4) high suction tensiometers record increasing pressures when plunged dry in free water. The findings provide useful guidance to researchers using these devices for suction measurement. © 2012 IEEE. |
Persistent Identifier | http://hdl.handle.net/10722/194458 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Lourenço, SDN | - |
dc.contributor.author | Gallipoli, D | - |
dc.contributor.author | Toll, DG | - |
dc.contributor.author | Augarde, CE | - |
dc.contributor.author | Evans, FD | - |
dc.date.accessioned | 2014-01-30T03:32:37Z | - |
dc.date.available | 2014-01-30T03:32:37Z | - |
dc.date.issued | 2012 | - |
dc.identifier.citation | 5th Asia-Pacific Conference on Unsaturated Soils 2012, 2012, v. 1, p. 385-391 | - |
dc.identifier.uri | http://hdl.handle.net/10722/194458 | - |
dc.description.abstract | In high suction tensiometers, cavitation occurs when the tension in the water inside the device exceeds a certain threshold. The threshold corresponds to the formation of air bubbles either by air entry into or formation within the porous stone or reservoir. The standard procedure to inhibit air bubble formation in the device is to flood under vacuum and apply water at high pressure to force any remaining air bubbles to dissolve. A series of factors are, however, likely to influence the cavitation behaviour of high suction tensiometers. This paper demonstrates that: (1) there are no precursory phenomena leading to cavitation, (2) higher pressures are not essential for saturation, (3) cooler temperatures and tensiometer usage decrease the pore water pressure at cavitation, and (4) high suction tensiometers record increasing pressures when plunged dry in free water. The findings provide useful guidance to researchers using these devices for suction measurement. © 2012 IEEE. | - |
dc.language | eng | - |
dc.relation.ispartof | 5th Asia-Pacific Conference on Unsaturated Soils 2012 | - |
dc.title | Cavitation in high suction tensiometers: Effect of temperature, time of use, and stone drying | - |
dc.type | Conference_Paper | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.scopus | eid_2-s2.0-84869858539 | - |
dc.identifier.volume | 1 | - |
dc.identifier.spage | 385 | - |
dc.identifier.epage | 391 | - |