File Download

There are no files associated with this item.

Supplementary

Conference Paper: Simultaneous multislice EPI reconstruction by incorporating split slice-GRAPPA with slice-dependent 2D Nyquist ghost correction (PEC-SP-SG)

TitleSimultaneous multislice EPI reconstruction by incorporating split slice-GRAPPA with slice-dependent 2D Nyquist ghost correction (PEC-SP-SG)
Authors
Issue Date2019
PublisherInternational Society for Magnetic Resonance in Medicine.
Citation
The 27th Annual Meeting & Exhibition of International Society for Magnetic Resonance in Medicine (ISMRM), Montreal, Canada, 11-16 May 2019, abstract no. p4613 How to Cite?
AbstractSimultaneous multislice (SMS) EPI reconstruction is challenging due to slice-dependent 2D phase differences between opposite polarities, which is collapsed across slices. Additionally, slice leakage is one major concern in some applications including diffusion and functional MRI. The proposed SMS EPI reconstruction incorporates phase error correction with split slice-GRAPPA (PEC-SP-SG), and was evaluated using simulation, phantom and in vivo experiments. Results show that the proposed approach can offer a robust SMS EPI reconstruction with slice-dependent 2D Nyquist ghost correction, and provide a balance between slice leakage and in-plane artifacts.
DescriptionDigital Poster Session -Acquisition, Reconstruction & Analysis: Artifacts, Implants & Corrections - Abstract #4613
Persistent Identifierhttp://hdl.handle.net/10722/275276

 

DC FieldValueLanguage
dc.contributor.authorLiu, Y-
dc.contributor.authorBarth, M-
dc.contributor.authorLyu, M-
dc.contributor.authorWu, EX-
dc.date.accessioned2019-09-10T02:39:15Z-
dc.date.available2019-09-10T02:39:15Z-
dc.date.issued2019-
dc.identifier.citationThe 27th Annual Meeting & Exhibition of International Society for Magnetic Resonance in Medicine (ISMRM), Montreal, Canada, 11-16 May 2019, abstract no. p4613-
dc.identifier.urihttp://hdl.handle.net/10722/275276-
dc.descriptionDigital Poster Session -Acquisition, Reconstruction & Analysis: Artifacts, Implants & Corrections - Abstract #4613-
dc.description.abstractSimultaneous multislice (SMS) EPI reconstruction is challenging due to slice-dependent 2D phase differences between opposite polarities, which is collapsed across slices. Additionally, slice leakage is one major concern in some applications including diffusion and functional MRI. The proposed SMS EPI reconstruction incorporates phase error correction with split slice-GRAPPA (PEC-SP-SG), and was evaluated using simulation, phantom and in vivo experiments. Results show that the proposed approach can offer a robust SMS EPI reconstruction with slice-dependent 2D Nyquist ghost correction, and provide a balance between slice leakage and in-plane artifacts.-
dc.languageeng-
dc.publisherInternational Society for Magnetic Resonance in Medicine.-
dc.relation.ispartofISMRM (International Society for Magnetic Resonance in Medicine) 27th Annual Meeting, 2019-
dc.titleSimultaneous multislice EPI reconstruction by incorporating split slice-GRAPPA with slice-dependent 2D Nyquist ghost correction (PEC-SP-SG)-
dc.typeConference_Paper-
dc.identifier.emailLiu, Y: loyalliu@hku.hk-
dc.identifier.emailWu, EX: ewu@eee.hku.hk-
dc.identifier.authorityWu, EX=rp00193-
dc.identifier.hkuros304420-
dc.identifier.hkuros307723-
dc.identifier.spagep4613-
dc.identifier.epagep4613-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats