File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: The Accuracy of Zygomatic Implant Placement Assisted by Dynamic Computer-Aided Surgery: A Systematic Review and Meta-Analysis

TitleThe Accuracy of Zygomatic Implant Placement Assisted by Dynamic Computer-Aided Surgery: A Systematic Review and Meta-Analysis
Authors
Keywordscomputer-aided surgery
edentulous
guided surgery
navigation
zygomatic implant
Issue Date2023
Citation
Journal of Clinical Medicine, 2023, v. 12, n. 16, article no. 5418 How to Cite?
AbstractPurpose: The present systematic review aimed to investigate the accuracy of zygomatic implant (ZI) placement using dynamic computer-aided surgery (d-CAIS), static computer-aided surgery (s-CAIS), and a free-hand approach in patients with severe atrophic edentulous maxilla and/or deficient maxilla. Methods: Electronic and manual literature searches until May 2023 were performed in the PubMed/Medline, Scopus, Cochrane Library, and Web of Science databases. Clinical trials and cadaver studies were selected. The primary outcome was planned/placed deviation. Secondary outcomes were to evaluate the survival of ZI and surgical complications. Random-effects meta-analyses were conducted and meta-regression was utilized to compare fiducial registration amounts for d-CAIS and the different designs of s-CAIS. Results: A total of 14 studies with 511 ZIs were included (Nobel Biocare: 274, Southern Implant: 42, SIN Implant: 16, non-mentioned: 179). The pooled mean ZI deviations from the d-CAIS group were 1.81 mm (95% CI: 1.34–2.29) at the entry point and 2.95 mm (95% CI: 1.66–4.24) at the apex point, and angular deviations were 3.49 degrees (95% CI: 2.04–4.93). The pooled mean ZI deviations from the s-CAIS group were 1.19 mm (95% CI: 0.83–1.54) at the entry point and 1.80 mm (95% CI: 1.10–2.50) at the apex point, and angular deviations were 2.15 degrees (95% CI: 1.43–2.88). The pooled mean ZI deviations from the free-hand group were 2.04 mm (95% CI: 1.69–2.39) at the entry point and 3.23 mm (95% CI: 2.34–4.12) at the apex point, and angular deviations were 4.92 degrees (95% CI: 3.86–5.98). There was strong evidence of differences in the average entry, apex, and angular deviation between the navigation, surgical guide, and free-hand groups (p < 0.01). A significant inverse correlation was observed between the number of fiducial screws and the planned/placed deviation regarding entry, apex, and angular measurements. Conclusion: Using d-CAIS and modified s-CAIS for ZI surgery has shown clinically acceptable outcomes regarding average entry, apex, and angular deviations. The maximal deviation values were predominantly observed in the conventional s-CAIS. Surgeons should be mindful of potential deviations and complications regardless of the decision making in different guide approaches.
Persistent Identifierhttp://hdl.handle.net/10722/354292
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorFan, Shengchi-
dc.contributor.authorSáenz-Ravello, Gustavo-
dc.contributor.authorDiaz, Leonardo-
dc.contributor.authorWu, Yiqun-
dc.contributor.authorDavó, Rubén-
dc.contributor.authorWang, Feng-
dc.contributor.authorMagic, Marko-
dc.contributor.authorAl-Nawas, Bilal-
dc.contributor.authorKämmerer, Peer W.-
dc.date.accessioned2025-02-07T08:47:42Z-
dc.date.available2025-02-07T08:47:42Z-
dc.date.issued2023-
dc.identifier.citationJournal of Clinical Medicine, 2023, v. 12, n. 16, article no. 5418-
dc.identifier.urihttp://hdl.handle.net/10722/354292-
dc.description.abstractPurpose: The present systematic review aimed to investigate the accuracy of zygomatic implant (ZI) placement using dynamic computer-aided surgery (d-CAIS), static computer-aided surgery (s-CAIS), and a free-hand approach in patients with severe atrophic edentulous maxilla and/or deficient maxilla. Methods: Electronic and manual literature searches until May 2023 were performed in the PubMed/Medline, Scopus, Cochrane Library, and Web of Science databases. Clinical trials and cadaver studies were selected. The primary outcome was planned/placed deviation. Secondary outcomes were to evaluate the survival of ZI and surgical complications. Random-effects meta-analyses were conducted and meta-regression was utilized to compare fiducial registration amounts for d-CAIS and the different designs of s-CAIS. Results: A total of 14 studies with 511 ZIs were included (Nobel Biocare: 274, Southern Implant: 42, SIN Implant: 16, non-mentioned: 179). The pooled mean ZI deviations from the d-CAIS group were 1.81 mm (95% CI: 1.34–2.29) at the entry point and 2.95 mm (95% CI: 1.66–4.24) at the apex point, and angular deviations were 3.49 degrees (95% CI: 2.04–4.93). The pooled mean ZI deviations from the s-CAIS group were 1.19 mm (95% CI: 0.83–1.54) at the entry point and 1.80 mm (95% CI: 1.10–2.50) at the apex point, and angular deviations were 2.15 degrees (95% CI: 1.43–2.88). The pooled mean ZI deviations from the free-hand group were 2.04 mm (95% CI: 1.69–2.39) at the entry point and 3.23 mm (95% CI: 2.34–4.12) at the apex point, and angular deviations were 4.92 degrees (95% CI: 3.86–5.98). There was strong evidence of differences in the average entry, apex, and angular deviation between the navigation, surgical guide, and free-hand groups (p < 0.01). A significant inverse correlation was observed between the number of fiducial screws and the planned/placed deviation regarding entry, apex, and angular measurements. Conclusion: Using d-CAIS and modified s-CAIS for ZI surgery has shown clinically acceptable outcomes regarding average entry, apex, and angular deviations. The maximal deviation values were predominantly observed in the conventional s-CAIS. Surgeons should be mindful of potential deviations and complications regardless of the decision making in different guide approaches.-
dc.languageeng-
dc.relation.ispartofJournal of Clinical Medicine-
dc.subjectcomputer-aided surgery-
dc.subjectedentulous-
dc.subjectguided surgery-
dc.subjectnavigation-
dc.subjectzygomatic implant-
dc.titleThe Accuracy of Zygomatic Implant Placement Assisted by Dynamic Computer-Aided Surgery: A Systematic Review and Meta-Analysis-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.3390/jcm12165418-
dc.identifier.scopuseid_2-s2.0-85169085945-
dc.identifier.volume12-
dc.identifier.issue16-
dc.identifier.spagearticle no. 5418-
dc.identifier.epagearticle no. 5418-
dc.identifier.eissn2077-0383-
dc.identifier.isiWOS:001056607100001-

Export via OAI-PMH Interface in XML Formats


OR