File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Geometric and dosimetric verification of a recurrent neural network algorithm to compensate for respiratory motion using an articulated robotic couch

TitleGeometric and dosimetric verification of a recurrent neural network algorithm to compensate for respiratory motion using an articulated robotic couch
Authors
KeywordsMotion compensation
RNN
Robotic couch
Issue Date2021
Citation
Journal of the Korean Physical Society, 2021, v. 78, n. 1, p. 64-72 How to Cite?
AbstractThe purpose of this study is to evaluate the performance of a recurrent neural network (RNN)-based prediction algorithm to compensate for respiratory movement using an articulated robotic couch system. A prototype of a real-time respiratory motion compensation couch was built using an optical 3D motion tracking system and a six-degree-of-freedom-articulated robotic system. To compensate for the system latency from motion detection to re-positioning of the system, RNN and double exponential smoothing (ES2) prediction algorithms were applied. Three aspects of performance were evaluated, simulation and experiments for geometric and dosimetric evaluations, using data from three liver and three lung patients who underwent stereotactic body radiotherapy. Overall, the RNN algorithm showed better geometric and dosimetric results than the other approaches. In simulation tests, RNN showed 82% average improvement ratio, compared with non-predicted results. In the geometric evaluation, RNN only showed average FWHM broadening of 1.5 mm, compared with the static case. In the dosimetric evaluation, RNN showed average gamma passing rates of 97.4 ± 1.0%, 89.0 ± 2.4% under the 3%/3 mm, 2%/2 mm respectively. It may be technically feasible to use the RNN prediction algorithm to compensate for respiratory motion with an articulated robotic couch system. The RNN algorithm could be widely used for motion compensation in patients undergoing radiotherapy.
Persistent Identifierhttp://hdl.handle.net/10722/345811
ISSN
2023 Impact Factor: 0.8
2023 SCImago Journal Rankings: 0.213

 

DC FieldValueLanguage
dc.contributor.authorLee, Minsik-
dc.contributor.authorCho, Min Seok-
dc.contributor.authorLee, Hoyeon-
dc.contributor.authorJeong, Chiyoung-
dc.contributor.authorKwak, Jungwon-
dc.contributor.authorJung, Jinhong-
dc.contributor.authorKim, Su Ssan-
dc.contributor.authorYoon, Sang Min-
dc.contributor.authorSong, Si Yeol-
dc.contributor.authorLee, Sang wook-
dc.contributor.authorKim, Jong Hoon-
dc.contributor.authorChoi, Eun Kyung-
dc.contributor.authorCho, Seungryong-
dc.contributor.authorCho, Byungchul-
dc.date.accessioned2024-09-01T10:59:51Z-
dc.date.available2024-09-01T10:59:51Z-
dc.date.issued2021-
dc.identifier.citationJournal of the Korean Physical Society, 2021, v. 78, n. 1, p. 64-72-
dc.identifier.issn0374-4884-
dc.identifier.urihttp://hdl.handle.net/10722/345811-
dc.description.abstractThe purpose of this study is to evaluate the performance of a recurrent neural network (RNN)-based prediction algorithm to compensate for respiratory movement using an articulated robotic couch system. A prototype of a real-time respiratory motion compensation couch was built using an optical 3D motion tracking system and a six-degree-of-freedom-articulated robotic system. To compensate for the system latency from motion detection to re-positioning of the system, RNN and double exponential smoothing (ES2) prediction algorithms were applied. Three aspects of performance were evaluated, simulation and experiments for geometric and dosimetric evaluations, using data from three liver and three lung patients who underwent stereotactic body radiotherapy. Overall, the RNN algorithm showed better geometric and dosimetric results than the other approaches. In simulation tests, RNN showed 82% average improvement ratio, compared with non-predicted results. In the geometric evaluation, RNN only showed average FWHM broadening of 1.5 mm, compared with the static case. In the dosimetric evaluation, RNN showed average gamma passing rates of 97.4 ± 1.0%, 89.0 ± 2.4% under the 3%/3 mm, 2%/2 mm respectively. It may be technically feasible to use the RNN prediction algorithm to compensate for respiratory motion with an articulated robotic couch system. The RNN algorithm could be widely used for motion compensation in patients undergoing radiotherapy.-
dc.languageeng-
dc.relation.ispartofJournal of the Korean Physical Society-
dc.subjectMotion compensation-
dc.subjectRNN-
dc.subjectRobotic couch-
dc.titleGeometric and dosimetric verification of a recurrent neural network algorithm to compensate for respiratory motion using an articulated robotic couch-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-