File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Fluoroscopic Frameless Computer-Assisted Navigation for Transsphenoidal Surgery: A Clinical Assessment of Accuracy in Spatial Position and Trajectory

TitleFluoroscopic Frameless Computer-Assisted Navigation for Transsphenoidal Surgery: A Clinical Assessment of Accuracy in Spatial Position and Trajectory
Authors
KeywordsAccuracy
Fluoroscopic navigation
Transsphenoidal surgery
Issue Date2004
Citation
Minimally Invasive Neurosurgery, 2004, v. 47, n. 1, p. 29-31 How to Cite?
AbstractObjective: Fluoroscopic navigation technique in transsphenoidal surgery on the one hand provides multi-planar (antero-posterior and lateral views) navigation while on the other hand it does not need any preoperative preparation. To assess the clinical accuracy of the above technique in the transsphenoidal surgery, we performed this study. Methods: 5 patients undergoing transsphenoidal surgery were assessed. Fluoroscopic images were compared with C-arm X-ray images for difference in spatial position and trajectory. Results: 26 sets of data were collected for analysis. Mean pointer tip difference was 0.6 mm. The 95% confidence interval was 1.3 mm. Mean trajectory difference was 1.2 degree. The 95 % confidence interval was 2.7 degree. Conclusion: With a good clinical accuracy, fluoroscopic navigation offers a distinct advantage to replace the traditional fluoroscopic approach for transsphenoidal surgery for its low X-ray exposure and multi-planar (antero-posterior and lateral views) navigation.
Persistent Identifierhttp://hdl.handle.net/10722/325104
ISSN
2013 Impact Factor: 1.143
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorWong, G. K.-
dc.contributor.authorPoon, W. S.-
dc.contributor.authorNg, S. C.-
dc.contributor.authorCheng, A. Y.-
dc.date.accessioned2023-02-27T07:29:45Z-
dc.date.available2023-02-27T07:29:45Z-
dc.date.issued2004-
dc.identifier.citationMinimally Invasive Neurosurgery, 2004, v. 47, n. 1, p. 29-31-
dc.identifier.issn0946-7211-
dc.identifier.urihttp://hdl.handle.net/10722/325104-
dc.description.abstractObjective: Fluoroscopic navigation technique in transsphenoidal surgery on the one hand provides multi-planar (antero-posterior and lateral views) navigation while on the other hand it does not need any preoperative preparation. To assess the clinical accuracy of the above technique in the transsphenoidal surgery, we performed this study. Methods: 5 patients undergoing transsphenoidal surgery were assessed. Fluoroscopic images were compared with C-arm X-ray images for difference in spatial position and trajectory. Results: 26 sets of data were collected for analysis. Mean pointer tip difference was 0.6 mm. The 95% confidence interval was 1.3 mm. Mean trajectory difference was 1.2 degree. The 95 % confidence interval was 2.7 degree. Conclusion: With a good clinical accuracy, fluoroscopic navigation offers a distinct advantage to replace the traditional fluoroscopic approach for transsphenoidal surgery for its low X-ray exposure and multi-planar (antero-posterior and lateral views) navigation.-
dc.languageeng-
dc.relation.ispartofMinimally Invasive Neurosurgery-
dc.subjectAccuracy-
dc.subjectFluoroscopic navigation-
dc.subjectTranssphenoidal surgery-
dc.titleFluoroscopic Frameless Computer-Assisted Navigation for Transsphenoidal Surgery: A Clinical Assessment of Accuracy in Spatial Position and Trajectory-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1055/s-2003-812535-
dc.identifier.pmid15100929-
dc.identifier.scopuseid_2-s2.0-2442571996-
dc.identifier.volume47-
dc.identifier.issue1-
dc.identifier.spage29-
dc.identifier.epage31-
dc.identifier.isiWOS:000221171000005-

Export via OAI-PMH Interface in XML Formats


OR