Article: Breakage of the radiopaque wire from the medullary tube during closed antegrade intramedullary nailing for femoral shaft fracture.
Title | Breakage of the radiopaque wire from the medullary tube during closed antegrade intramedullary nailing for femoral shaft fracture. |
---|---|
Authors | |
Keywords | Femur Fracture fixation, intramedullary Fractures, bone |
Issue Date | 2012 |
Citation | Journal of Orthopaedic Surgery, 2012, v. 20, n. 1, p. 118-120 How to Cite? |
Abstract | We report a complication of radiopaque wire breakage from the medullary tube during closed antegrade intramedullary nailing for a femoral shaft fracture. To avoid such complication, the medullary tube should be checked carefully for colour changes and surface defects, and tested for flexibility before each use. The medullary tube should also be replaced before 100 exposures to autoclaving. |
Persistent Identifier | http://hdl.handle.net/10722/309463 |
ISSN | 2023 Impact Factor: 1.3 2023 SCImago Journal Rankings: 0.557 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Chan, Ping Keung | - |
dc.contributor.author | Chan, Kwok Bill | - |
dc.contributor.author | Lui, Tun Hing | - |
dc.contributor.author | Ngai, Wai Kit | - |
dc.date.accessioned | 2021-12-29T07:02:29Z | - |
dc.date.available | 2021-12-29T07:02:29Z | - |
dc.date.issued | 2012 | - |
dc.identifier.citation | Journal of Orthopaedic Surgery, 2012, v. 20, n. 1, p. 118-120 | - |
dc.identifier.issn | 1022-5536 | - |
dc.identifier.uri | http://hdl.handle.net/10722/309463 | - |
dc.description.abstract | We report a complication of radiopaque wire breakage from the medullary tube during closed antegrade intramedullary nailing for a femoral shaft fracture. To avoid such complication, the medullary tube should be checked carefully for colour changes and surface defects, and tested for flexibility before each use. The medullary tube should also be replaced before 100 exposures to autoclaving. | - |
dc.language | eng | - |
dc.relation.ispartof | Journal of Orthopaedic Surgery | - |
dc.subject | Femur | - |
dc.subject | Fracture fixation, intramedullary | - |
dc.subject | Fractures, bone | - |
dc.title | Breakage of the radiopaque wire from the medullary tube during closed antegrade intramedullary nailing for femoral shaft fracture. | - |
dc.type | Article | - |
dc.description.nature | link_to_OA_fulltext | - |
dc.identifier.doi | 10.1177/230949901202000126 | - |
dc.identifier.pmid | 22535827 | - |
dc.identifier.scopus | eid_2-s2.0-84865113483 | - |
dc.identifier.volume | 20 | - |
dc.identifier.issue | 1 | - |
dc.identifier.spage | 118 | - |
dc.identifier.epage | 120 | - |
dc.identifier.isi | WOS:000209171000026 | - |