Article: Importance of the epiphyseal ring in OLIF stand-alone surgery: a biomechanical study on cadaveric spines

TitleImportance of the epiphyseal ring in OLIF stand-alone surgery: a biomechanical study on cadaveric spines
Authors
KeywordsEpiphyseal ring
Stand-alone surgery
Biomechanical model
OLIF
Endplate collapse
Issue Date2021
Citation
European Spine Journal, 2021, v. 30, n. 1, p. 79-87 How to Cite?
AbstractPurposes: To explore the function of endplate epiphyseal ring in OLIF stand-alone surgery using a biomechanical model to reduce the complications of endplate collapse and cage subsidence. Methods: In total, 24 human cadaveric lumbar function units (L1–2 and L3–4 segments) were randomly assigned to two groups. The first group was implanted with long fusion cages which engaged with both inner and outer regions of epiphyseal ring (Complete Span-Epiphyseal Ring, CSER). Those engaged with only the inner half of epiphyseal ring were the second group (Half Span-Epiphyseal Ring, HSER). Each group was divided into two subgroups [higher cage-height (HH) and normal cage-height (NH)]. Specimens were fixed in testing cups and compressed at approximately 2.5 mm/s, until the first sign of structural failure. Trabecular structural damage was analyzed by Micro-CT, as well as the difference of bone volume fraction (BV/TV), trabecular thickness (Tb.Th) et al. in different regions. Results: Endplate collapse was mainly evident in the inner region of epiphyseal ring, where trabecular injury of sub-endplate bone was most concentrated. Endplate collapse incidence was significantly higher in HSER than CSER specimens (P = 0.017). A structural failure occurred at a lower force in HSER (1.41 ± 0.34 KN) compared with CSER (2.44 ± 0.59 KN). HH subgroups failed at a lower average force than NH subgroups. Micro-CT results showed a more extensive trabecular fracture in HSER specimens compared to CSER specimens, especially in HH subgroup. Conclusions: Endplate collapse is more likely to occur with short half span cages than complete span cages, and taller cages compared with normal height cages. During OLIF surgery, we should choose cages matching intervertebral disc space height and place the cages spanning over the whole epiphyseal ring to improve support strength.
Persistent Identifierhttp://hdl.handle.net/10722/301859
ISSN
2023 Impact Factor: 2.6
2023 SCImago Journal Rankings: 1.042
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorZhang, Xuyang-
dc.contributor.authorWu, Hao-
dc.contributor.authorChen, Yilei-
dc.contributor.authorLiu, Junhui-
dc.contributor.authorChen, Jian-
dc.contributor.authorZhang, Teng-
dc.contributor.authorZhou, Zhao Feng-
dc.contributor.authorFan, Shunwu-
dc.contributor.authorDolan, Patricia-
dc.contributor.authorAdams, Michael Anthony-
dc.contributor.authorZhao, Fengdong-
dc.date.accessioned2021-08-19T02:20:53Z-
dc.date.available2021-08-19T02:20:53Z-
dc.date.issued2021-
dc.identifier.citationEuropean Spine Journal, 2021, v. 30, n. 1, p. 79-87-
dc.identifier.issn0940-6719-
dc.identifier.urihttp://hdl.handle.net/10722/301859-
dc.description.abstractPurposes: To explore the function of endplate epiphyseal ring in OLIF stand-alone surgery using a biomechanical model to reduce the complications of endplate collapse and cage subsidence. Methods: In total, 24 human cadaveric lumbar function units (L1–2 and L3–4 segments) were randomly assigned to two groups. The first group was implanted with long fusion cages which engaged with both inner and outer regions of epiphyseal ring (Complete Span-Epiphyseal Ring, CSER). Those engaged with only the inner half of epiphyseal ring were the second group (Half Span-Epiphyseal Ring, HSER). Each group was divided into two subgroups [higher cage-height (HH) and normal cage-height (NH)]. Specimens were fixed in testing cups and compressed at approximately 2.5 mm/s, until the first sign of structural failure. Trabecular structural damage was analyzed by Micro-CT, as well as the difference of bone volume fraction (BV/TV), trabecular thickness (Tb.Th) et al. in different regions. Results: Endplate collapse was mainly evident in the inner region of epiphyseal ring, where trabecular injury of sub-endplate bone was most concentrated. Endplate collapse incidence was significantly higher in HSER than CSER specimens (P = 0.017). A structural failure occurred at a lower force in HSER (1.41 ± 0.34 KN) compared with CSER (2.44 ± 0.59 KN). HH subgroups failed at a lower average force than NH subgroups. Micro-CT results showed a more extensive trabecular fracture in HSER specimens compared to CSER specimens, especially in HH subgroup. Conclusions: Endplate collapse is more likely to occur with short half span cages than complete span cages, and taller cages compared with normal height cages. During OLIF surgery, we should choose cages matching intervertebral disc space height and place the cages spanning over the whole epiphyseal ring to improve support strength.-
dc.languageeng-
dc.relation.ispartofEuropean Spine Journal-
dc.subjectEpiphyseal ring-
dc.subjectStand-alone surgery-
dc.subjectBiomechanical model-
dc.subjectOLIF-
dc.subjectEndplate collapse-
dc.titleImportance of the epiphyseal ring in OLIF stand-alone surgery: a biomechanical study on cadaveric spines-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1007/s00586-020-06667-2-
dc.identifier.pmid33226482-
dc.identifier.scopuseid_2-s2.0-85096436632-
dc.identifier.volume30-
dc.identifier.issue1-
dc.identifier.spage79-
dc.identifier.epage87-
dc.identifier.eissn1432-0932-
dc.identifier.isiWOS:000591963800001-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats