Article: Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes

TitleCorneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes
Authors
KeywordsKeratoconus
Corneal topography
Topography
Issue Date2013
Citation
Clinical and Experimental Ophthalmology, 2013, v. 41, n. 8, p. 735-745 How to Cite?
AbstractBackground: To compare corneal thickness and corneal elevation using swept source optical coherence tomography and slit scanning topography. Design: Prospective study. Participants: 41 normal and 46 keratoconus subjects. Methods: All eyes were imaged using swept source optical coherence tomography and slit scanning tomography during the same visit. Mean corneal thickness and best-fit sphere measurements were compared between the instruments. Main outcome measures: Agreement of measurements between swept source optical coherence tomography and scanning slit topography was analyzed. Intra-rater reproducibility coefficient and intraclass correlation coefficient were evaluated. Results: In normal eyes, central corneal thickness measured by swept source optical coherence tomography was thinner compared with slit scanning topography (p<0.0001) and ultrasound pachymetry (p=<.0001). Ultrasound pachymetry readings had better 95% limits of agreement with swept source optical coherence tomography than slit scanning topography. In keratoconus eyes, central corneal thickness was thinner on swept source optical coherence tomography than slit scanning topography (p=0.081) and ultrasound pachymetry (p=0.001). There were significant differences between thinnest corneal thickness, and, anterior and posterior best-fit sphere measurements between both instruments (p<0.05 for all). Overall, reproducibility coefficients and intraclass correlation coefficients were significantly better with swept source optical coherence tomography for measurement of central corneal thickness, anterior best-fit sphere and, posterior best-fit sphere (all p<0.001). Conclusions: Corneal thickness and elevation measurements were significantly different between swept source optical coherence tomography and slit scanning topography. With better reproducibility coefficients and intraclass correlation coefficients, swept source optical coherence tomography may provide a reliable alternative for measurement of corneal parameters. © 2013 Royal Australian and New Zealand College of Ophthalmologists.
Persistent Identifierhttp://hdl.handle.net/10722/298055
ISSN
2023 Impact Factor: 4.9
2023 SCImago Journal Rankings: 1.368
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorJhanji, Vishal-
dc.contributor.authorYang, Bingzhi-
dc.contributor.authorYu, Marco-
dc.contributor.authorYe, Cong-
dc.contributor.authorLeung, Christopher K.S.-
dc.date.accessioned2021-04-08T03:07:33Z-
dc.date.available2021-04-08T03:07:33Z-
dc.date.issued2013-
dc.identifier.citationClinical and Experimental Ophthalmology, 2013, v. 41, n. 8, p. 735-745-
dc.identifier.issn1442-6404-
dc.identifier.urihttp://hdl.handle.net/10722/298055-
dc.description.abstractBackground: To compare corneal thickness and corneal elevation using swept source optical coherence tomography and slit scanning topography. Design: Prospective study. Participants: 41 normal and 46 keratoconus subjects. Methods: All eyes were imaged using swept source optical coherence tomography and slit scanning tomography during the same visit. Mean corneal thickness and best-fit sphere measurements were compared between the instruments. Main outcome measures: Agreement of measurements between swept source optical coherence tomography and scanning slit topography was analyzed. Intra-rater reproducibility coefficient and intraclass correlation coefficient were evaluated. Results: In normal eyes, central corneal thickness measured by swept source optical coherence tomography was thinner compared with slit scanning topography (p<0.0001) and ultrasound pachymetry (p=<.0001). Ultrasound pachymetry readings had better 95% limits of agreement with swept source optical coherence tomography than slit scanning topography. In keratoconus eyes, central corneal thickness was thinner on swept source optical coherence tomography than slit scanning topography (p=0.081) and ultrasound pachymetry (p=0.001). There were significant differences between thinnest corneal thickness, and, anterior and posterior best-fit sphere measurements between both instruments (p<0.05 for all). Overall, reproducibility coefficients and intraclass correlation coefficients were significantly better with swept source optical coherence tomography for measurement of central corneal thickness, anterior best-fit sphere and, posterior best-fit sphere (all p<0.001). Conclusions: Corneal thickness and elevation measurements were significantly different between swept source optical coherence tomography and slit scanning topography. With better reproducibility coefficients and intraclass correlation coefficients, swept source optical coherence tomography may provide a reliable alternative for measurement of corneal parameters. © 2013 Royal Australian and New Zealand College of Ophthalmologists.-
dc.languageeng-
dc.relation.ispartofClinical and Experimental Ophthalmology-
dc.subjectKeratoconus-
dc.subjectCorneal topography-
dc.subjectTopography-
dc.titleCorneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1111/ceo.12113-
dc.identifier.pmid23566209-
dc.identifier.scopuseid_2-s2.0-84886303135-
dc.identifier.volume41-
dc.identifier.issue8-
dc.identifier.spage735-
dc.identifier.epage745-
dc.identifier.eissn1442-9071-
dc.identifier.isiWOS:000326027600004-
dc.identifier.issnl1442-6404-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats