Article: Comparative Study of Central Corneal Thickness Measurement with Slit-Lamp Optical Coherence Tomography and Visante Optical Coherence Tomography
Title | Comparative Study of Central Corneal Thickness Measurement with Slit-Lamp Optical Coherence Tomography and Visante Optical Coherence Tomography |
---|---|
Authors | |
Issue Date | 2008 |
Citation | Ophthalmology, 2008, v. 115, n. 5, p. 796-801.e2 How to Cite? |
Abstract | Purpose: To evaluate the repeatability and reproducibility of central corneal thickness (CCT) measurements obtained by 2 anterior segment optical coherence tomography (OCT) imaging systems and to examine their agreements with ultrasound pachymetry. Design: Observational cross-sectional study. Participants: Fifty eyes from 50 healthy normal subjects were recruited. Methods: In one randomly selected eye in each subject, CCT was measured by slit-lamp OCT (SLOCT), Visante OCT, and ultrasound pachymetry. For anterior segment OCT measurements, both automatic and manual CCTs were obtained. Twenty-five of the 50 subjects were invited for 2 more visits within a week to evaluate repeatability and reproducibility of CCT measurement. Main Outcome Measures: Central corneal thickness measurement obtained by the 3 methods and their agreements. Intrasession and intersession within-subject standard deviation (S ), precision (1.96×S ), coefficient of variation (CV ) (100×S /overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility. Results: Good repeatability and reproducibility were found for both automatic and manual CCT measurements obtained by SLOCT and Visante OCT. For intrasession repeatability, CV and ICC values ranged between 0.9% and 1.2% and 0.96 and 0.98, respectively. For intersession reproducibility, the respective CV and ICC values ranged between 1.2% and 1.4% and 0.94 and 0.96. Although no significant difference was found between automatic/manual SLOCT measurements and ultrasound pachymetry, automatic Visante OCT CCT (535.7±30.2 μm) was significantly less than CCT with ultrasound pachymetry (550.3±31.14 μm) (P<0.001). In contrast, manual Visante OCT measurement (558.8±32.8 μm) was slightly higher than ultrasound pachymetry (P<0.001). Nevertheless, SLOCT and Visante OCT measurement of CCT had 95% limits of agreement comparable to that of ultrasound pachymetry. The best agreement was observed in the manual SLOCT measurement (95% limits of agreement between -15.5 and 11.7 μm). Conclusions: Both SLOCT and Visante OCT automatic and manual CCT measurements were reliable and showed comparable agreement with ultrasound pachymetry. Although the 2 anterior segment OCT imaging systems have similar design and working principles, clinicians should be aware of the differences in CCT measurement between the 2 anterior segment OCTs. © 2008 American Academy of Ophthalmology. w w w w w w |
Persistent Identifier | http://hdl.handle.net/10722/297964 |
ISSN | 2023 Impact Factor: 13.1 2023 SCImago Journal Rankings: 4.642 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Li, Haitao | - |
dc.contributor.author | Leung, Christopher Kai Shun | - |
dc.contributor.author | Wong, Lee | - |
dc.contributor.author | Cheung, Carol Yim Lui | - |
dc.contributor.author | Pang, Chi Pui | - |
dc.contributor.author | Weinreb, Robert Neal | - |
dc.contributor.author | Lam, Dennis Shun Chiu | - |
dc.date.accessioned | 2021-04-08T03:07:21Z | - |
dc.date.available | 2021-04-08T03:07:21Z | - |
dc.date.issued | 2008 | - |
dc.identifier.citation | Ophthalmology, 2008, v. 115, n. 5, p. 796-801.e2 | - |
dc.identifier.issn | 0161-6420 | - |
dc.identifier.uri | http://hdl.handle.net/10722/297964 | - |
dc.description.abstract | Purpose: To evaluate the repeatability and reproducibility of central corneal thickness (CCT) measurements obtained by 2 anterior segment optical coherence tomography (OCT) imaging systems and to examine their agreements with ultrasound pachymetry. Design: Observational cross-sectional study. Participants: Fifty eyes from 50 healthy normal subjects were recruited. Methods: In one randomly selected eye in each subject, CCT was measured by slit-lamp OCT (SLOCT), Visante OCT, and ultrasound pachymetry. For anterior segment OCT measurements, both automatic and manual CCTs were obtained. Twenty-five of the 50 subjects were invited for 2 more visits within a week to evaluate repeatability and reproducibility of CCT measurement. Main Outcome Measures: Central corneal thickness measurement obtained by the 3 methods and their agreements. Intrasession and intersession within-subject standard deviation (S ), precision (1.96×S ), coefficient of variation (CV ) (100×S /overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility. Results: Good repeatability and reproducibility were found for both automatic and manual CCT measurements obtained by SLOCT and Visante OCT. For intrasession repeatability, CV and ICC values ranged between 0.9% and 1.2% and 0.96 and 0.98, respectively. For intersession reproducibility, the respective CV and ICC values ranged between 1.2% and 1.4% and 0.94 and 0.96. Although no significant difference was found between automatic/manual SLOCT measurements and ultrasound pachymetry, automatic Visante OCT CCT (535.7±30.2 μm) was significantly less than CCT with ultrasound pachymetry (550.3±31.14 μm) (P<0.001). In contrast, manual Visante OCT measurement (558.8±32.8 μm) was slightly higher than ultrasound pachymetry (P<0.001). Nevertheless, SLOCT and Visante OCT measurement of CCT had 95% limits of agreement comparable to that of ultrasound pachymetry. The best agreement was observed in the manual SLOCT measurement (95% limits of agreement between -15.5 and 11.7 μm). Conclusions: Both SLOCT and Visante OCT automatic and manual CCT measurements were reliable and showed comparable agreement with ultrasound pachymetry. Although the 2 anterior segment OCT imaging systems have similar design and working principles, clinicians should be aware of the differences in CCT measurement between the 2 anterior segment OCTs. © 2008 American Academy of Ophthalmology. w w w w w w | - |
dc.language | eng | - |
dc.relation.ispartof | Ophthalmology | - |
dc.title | Comparative Study of Central Corneal Thickness Measurement with Slit-Lamp Optical Coherence Tomography and Visante Optical Coherence Tomography | - |
dc.type | Article | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1016/j.ophtha.2007.07.006 | - |
dc.identifier.pmid | 17916376 | - |
dc.identifier.scopus | eid_2-s2.0-42649140555 | - |
dc.identifier.volume | 115 | - |
dc.identifier.issue | 5 | - |
dc.identifier.spage | 796 | - |
dc.identifier.epage | 801.e2 | - |
dc.identifier.isi | WOS:000255614400007 | - |
dc.identifier.issnl | 0161-6420 | - |