Article: Effect of Spacers on the Bipolar Electrostatic Charge Properties of Metered Dose Inhaler Aerosols—A Case Study With Tilade®

TitleEffect of Spacers on the Bipolar Electrostatic Charge Properties of Metered Dose Inhaler Aerosols—A Case Study With Tilade®
Authors
Keywordsstatic charge
pMDI (pressurized metered dose inhaler)
aerosol
pulmonary
dispersion
Issue Date2017
PublisherJohn Wiley & Sons, Inc. The Journal's web site is located at http://www.acs.org
Citation
Journal of Pharmaceutical Sciences, 2017, v. 106 n. 6, p. 1553-1559 How to Cite?
AbstractAerosols emitted from metered dose inhalers (MDIs) are generally electrically charged and bipolar in nature. Although a spacer can effectively dampen the charge magnitude of aerosols, the electrostatic interactions between the positively and negatively charged particles and the spacer have not yet been characterized separately. The Bipolar Charge Analyzer (BOLAR) was employed to investigate interactions between the spacer and the charged aerosols. Three individual actuations of Tilade (R) MDI were introduced without a spacer and through an antistatic AeroChamber Plus (R) Z Stat (R), an uncoated and a detergent-coated AeroChamber Plus (R) spacer into the BOLAR at 60 L/min. Charge and mass profiles were determined. The surface potential of spacers followed the order of uncoated > detergent-coated > antistatic spacer. The spacers had minimal impact on the positively charged particles but the charge magnitude of the negatively charged particles was in the opposite order as the spacer surface potential. The charge-to-mass ratio of particles had little alteration for all measurements. Negatively charged particles had a higher tendency to deposit on the spacer walls, possibly due to their higher abundance in the confined spacer volume. The bipolar data may prove useful for designing better MDIs and spacers and modelling lung deposition of charged aerosol particles. (C) 2017 American Pharmacists Association (R). Published by Elsevier Inc. All rights reserved.
Persistent Identifierhttp://hdl.handle.net/10722/297684
ISSN
2023 Impact Factor: 3.7
2023 SCImago Journal Rankings: 0.705
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorLeung, SSY-
dc.contributor.authorChiow, ACM-
dc.contributor.authorKwok, PCL-
dc.contributor.authorUkkonen, ARI-
dc.contributor.authorChan, HK-
dc.date.accessioned2021-03-23T04:20:15Z-
dc.date.available2021-03-23T04:20:15Z-
dc.date.issued2017-
dc.identifier.citationJournal of Pharmaceutical Sciences, 2017, v. 106 n. 6, p. 1553-1559-
dc.identifier.issn0022-3549-
dc.identifier.urihttp://hdl.handle.net/10722/297684-
dc.description.abstractAerosols emitted from metered dose inhalers (MDIs) are generally electrically charged and bipolar in nature. Although a spacer can effectively dampen the charge magnitude of aerosols, the electrostatic interactions between the positively and negatively charged particles and the spacer have not yet been characterized separately. The Bipolar Charge Analyzer (BOLAR) was employed to investigate interactions between the spacer and the charged aerosols. Three individual actuations of Tilade (R) MDI were introduced without a spacer and through an antistatic AeroChamber Plus (R) Z Stat (R), an uncoated and a detergent-coated AeroChamber Plus (R) spacer into the BOLAR at 60 L/min. Charge and mass profiles were determined. The surface potential of spacers followed the order of uncoated > detergent-coated > antistatic spacer. The spacers had minimal impact on the positively charged particles but the charge magnitude of the negatively charged particles was in the opposite order as the spacer surface potential. The charge-to-mass ratio of particles had little alteration for all measurements. Negatively charged particles had a higher tendency to deposit on the spacer walls, possibly due to their higher abundance in the confined spacer volume. The bipolar data may prove useful for designing better MDIs and spacers and modelling lung deposition of charged aerosol particles. (C) 2017 American Pharmacists Association (R). Published by Elsevier Inc. All rights reserved.-
dc.languageeng-
dc.publisherJohn Wiley & Sons, Inc. The Journal's web site is located at http://www.acs.org-
dc.relation.ispartofJournal of Pharmaceutical Sciences-
dc.rightsSubmitted (preprint) Version This is the pre-peer reviewed version of the following article: [FULL CITE], which has been published in final form at [Link to final article using the DOI]. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions. Accepted (peer-reviewed) Version This is the peer reviewed version of the following article: [FULL CITE], which has been published in final form at [Link to final article using the DOI]. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions.-
dc.subjectstatic charge-
dc.subjectpMDI (pressurized metered dose inhaler)-
dc.subjectaerosol-
dc.subjectpulmonary-
dc.subjectdispersion-
dc.titleEffect of Spacers on the Bipolar Electrostatic Charge Properties of Metered Dose Inhaler Aerosols—A Case Study With Tilade®-
dc.typeArticle-
dc.identifier.emailKwok, PCL: pclkwok@hku.hk-
dc.identifier.authorityKwok, PCL=rp01540-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1016/j.xphs.2017.02.014-
dc.identifier.pmid28238902-
dc.identifier.scopuseid_2-s2.0-85016450093-
dc.identifier.hkuros321757-
dc.identifier.volume106-
dc.identifier.issue6-
dc.identifier.spage1553-
dc.identifier.epage1559-
dc.identifier.isiWOS:000407602700012-
dc.publisher.placeUnited States-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats