Article: The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor
Title | The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor |
---|---|
Authors | |
Keywords | diagnostic techniques and procedures ultrasonography hemodynamics Doppler pediatrics resuscitation |
Issue Date | 2010 |
Citation | Critical Care Medicine, 2010, v. 38, n. 9, p. 1875-1881 How to Cite? |
Abstract | Objective: The Ultrasonic Cardiac Output Monitor is a noninvasive method of hemodynamic assessment and monitoring in critically ill patients. There are no published reference ranges for normal values in children for this device. This study aimed to establish normal ranges for cardiovascular indices measured using Ultrasonic Cardiac Output Monitor in children aged 0-12 yrs old and to assess interobserver reliability. Design: This was a population-based cross-sectional observational study. Setting: Schools and kindergartens in Hong Kong. Subjects: Chinese children aged up to 12 yrs old. Interventions: Two operators performed Ultrasonic Cardiac Output Monitor scans on each child together with standard oscillometric measurement of blood pressure and heart rate. Software intrinsic to the Ultrasonic Cardiac Output Monitor device produces values for stroke volume, cardiac output, and systemic vascular resistance. For each parameter, normal ranges were defined as lying between the 2.5th and 97.5th percentiles. Interobserver reliability was assessed with Bland-Altman plots, coefficients of variation, and intraclass correlation. Measurements and main results: A total of 1,197 Chinese children (55% boys) were scanned. Normal ranges of values for cardiac output, stroke volume, and systemic vascular resistance indices are presented. Interobserver reliability for Ultrasonic Cardiac Output Monitor was superior to that for standard blood pressure and heart rate measurement. Conclusions: This large study presents normal values for cardiovascular indices in children using the Ultrasonic Cardiac Output Monitor with good interobserver reliability. © 2010 by the Society of Critical Care Medicine and Lippincott Williams & Wilkins. |
Persistent Identifier | http://hdl.handle.net/10722/291985 |
ISSN | 2023 Impact Factor: 7.7 2023 SCImago Journal Rankings: 2.663 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Cattermole, Giles N. | - |
dc.contributor.author | Leung, P. Y.Mia | - |
dc.contributor.author | Mak, Paulina S.K. | - |
dc.contributor.author | Chan, Stewart S.W. | - |
dc.contributor.author | Graham, Colin A. | - |
dc.contributor.author | Rainer, Timothy H. | - |
dc.date.accessioned | 2020-11-17T14:55:31Z | - |
dc.date.available | 2020-11-17T14:55:31Z | - |
dc.date.issued | 2010 | - |
dc.identifier.citation | Critical Care Medicine, 2010, v. 38, n. 9, p. 1875-1881 | - |
dc.identifier.issn | 0090-3493 | - |
dc.identifier.uri | http://hdl.handle.net/10722/291985 | - |
dc.description.abstract | Objective: The Ultrasonic Cardiac Output Monitor is a noninvasive method of hemodynamic assessment and monitoring in critically ill patients. There are no published reference ranges for normal values in children for this device. This study aimed to establish normal ranges for cardiovascular indices measured using Ultrasonic Cardiac Output Monitor in children aged 0-12 yrs old and to assess interobserver reliability. Design: This was a population-based cross-sectional observational study. Setting: Schools and kindergartens in Hong Kong. Subjects: Chinese children aged up to 12 yrs old. Interventions: Two operators performed Ultrasonic Cardiac Output Monitor scans on each child together with standard oscillometric measurement of blood pressure and heart rate. Software intrinsic to the Ultrasonic Cardiac Output Monitor device produces values for stroke volume, cardiac output, and systemic vascular resistance. For each parameter, normal ranges were defined as lying between the 2.5th and 97.5th percentiles. Interobserver reliability was assessed with Bland-Altman plots, coefficients of variation, and intraclass correlation. Measurements and main results: A total of 1,197 Chinese children (55% boys) were scanned. Normal ranges of values for cardiac output, stroke volume, and systemic vascular resistance indices are presented. Interobserver reliability for Ultrasonic Cardiac Output Monitor was superior to that for standard blood pressure and heart rate measurement. Conclusions: This large study presents normal values for cardiovascular indices in children using the Ultrasonic Cardiac Output Monitor with good interobserver reliability. © 2010 by the Society of Critical Care Medicine and Lippincott Williams & Wilkins. | - |
dc.language | eng | - |
dc.relation.ispartof | Critical Care Medicine | - |
dc.subject | diagnostic techniques and procedures | - |
dc.subject | ultrasonography | - |
dc.subject | hemodynamics | - |
dc.subject | Doppler | - |
dc.subject | pediatrics | - |
dc.subject | resuscitation | - |
dc.title | The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor | - |
dc.type | Article | - |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1097/CCM.0b013e3181e8adee | - |
dc.identifier.pmid | 20562697 | - |
dc.identifier.scopus | eid_2-s2.0-77956231366 | - |
dc.identifier.volume | 38 | - |
dc.identifier.issue | 9 | - |
dc.identifier.spage | 1875 | - |
dc.identifier.epage | 1881 | - |
dc.identifier.eissn | 1530-0293 | - |
dc.identifier.isi | WOS:000281559500015 | - |
dc.identifier.issnl | 0090-3493 | - |