File Download
  Links for fulltext
     (May Require Subscription)
Supplementary

Article: The Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure

TitleThe Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure
Authors
KeywordsComplication
Deformed threads
Distraction failure mechanism
Magnetically controlled growing rod
Rod slippage
Issue Date2020
PublisherLippincott, Williams & Wilkins. The Journal's web site is located at http://journals.lww.com/spinejournal/pages/default.aspx
Citation
Spine, 2020, v. 45 n. 6, p. E346-E351 How to Cite?
AbstractSTUDY DESIGN: Case report OBJECTIVE.: To report a unique mechanical failure of magnetically controlled growing rods (MCGR) that is related to continuous rod slippage. SUMMARY OF BACKGROUND DATA: Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. This complication ultimately led to distraction failure and revision surgery. METHODS: A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale 'crooked rod' radiological sign was identified by angulation between the actuator and the extendable portion of the rod. RESULTS: Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads. CONCLUSIONS: Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions. LEVEL OF EVIDENCE: 4. This is a report of a mechanical failure with magnetically controlled growing rods resulting from continuous distraction attempts despite persistent rod slippage. The internal screw's threads were deformed and the magnetically controlled growing rods could no longer be lengthened. Distraction failure and rod exchange was necessary to proceed to further lengthening.
Persistent Identifierhttp://hdl.handle.net/10722/279699
ISSN
2021 Impact Factor: 3.241
2020 SCImago Journal Rankings: 1.657
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorCheung, JPY-
dc.contributor.authorZhang, T-
dc.contributor.authorBow, HYC-
dc.contributor.authorKwan, KYH-
dc.contributor.authorSze, KY-
dc.contributor.authorCheung, KMC-
dc.date.accessioned2019-12-09T06:44:36Z-
dc.date.available2019-12-09T06:44:36Z-
dc.date.issued2020-
dc.identifier.citationSpine, 2020, v. 45 n. 6, p. E346-E351-
dc.identifier.issn0362-2436-
dc.identifier.urihttp://hdl.handle.net/10722/279699-
dc.description.abstractSTUDY DESIGN: Case report OBJECTIVE.: To report a unique mechanical failure of magnetically controlled growing rods (MCGR) that is related to continuous rod slippage. SUMMARY OF BACKGROUND DATA: Despite the growing popularity of MCGR in the management of patients with early onset scoliosis, postoperative complications and reoperations are not uncommon. Unique complications or unplanned reoperations are observed in patients with MCGRs that are not seen in traditional growing rods. The complications include rod slippage, mismatch between targeted and achieved distraction length, metallosis and actuator pin fracture. We have identified an unreported failure mechanism whereby deformed threads occur in the internal distraction mechanism of the MCGR. This phenomenon may indicate increased MCGR internal screw friction and increased distraction resistance. This complication ultimately led to distraction failure and revision surgery. METHODS: A girl with early onset scoliosis (EOS) was treated by dual MCGRs. She developed proximal junctional kyphosis (PJK) and continuous rod slippage. Distractions in the out-patient clinic and under sedation did not achieve successful rod lengthening. A tell-tale 'crooked rod' radiological sign was identified by angulation between the actuator and the extendable portion of the rod. RESULTS: Rod exchange was performed and she is now 2 years after revision surgery with successful lengthening episodes. Upon review of the extracted MCGR, distraction was not possible even after rod removal and was only resumed after the screw was manually reinserted along its correct threads. CONCLUSIONS: Deformed threads is a complication that cannot be reverted without rod removal and exchange. Users should be aware of this potential failure mechanism and proceed to early rod exchange rather than attempting any further distractions. LEVEL OF EVIDENCE: 4. This is a report of a mechanical failure with magnetically controlled growing rods resulting from continuous distraction attempts despite persistent rod slippage. The internal screw's threads were deformed and the magnetically controlled growing rods could no longer be lengthened. Distraction failure and rod exchange was necessary to proceed to further lengthening.-
dc.languageeng-
dc.publisherLippincott, Williams & Wilkins. The Journal's web site is located at http://journals.lww.com/spinejournal/pages/default.aspx-
dc.relation.ispartofSpine-
dc.rightsThis is a non-final version of an article published in final form in Spine, 2020, v. 45 n. 6, p. E346-E351-
dc.subjectComplication-
dc.subjectDeformed threads-
dc.subjectDistraction failure mechanism-
dc.subjectMagnetically controlled growing rod-
dc.subjectRod slippage-
dc.titleThe Crooked Rod Sign: A New Radiological Sign to Detect Deformed Threads in the Distraction Mechanism of Magnetically Controlled Growing Rods and a Mode of Distraction Failure-
dc.typeArticle-
dc.identifier.emailCheung, JPY: cheungjp@hku.hk-
dc.identifier.emailZhang, T: tgzhang@hku.hk-
dc.identifier.emailBow, HYC: cbow@hku.hk-
dc.identifier.emailKwan, KYH: kyhkwan@hku.hk-
dc.identifier.emailSze, KY: kysze@hku.hk-
dc.identifier.emailCheung, KMC: cheungmc@hku.hk-
dc.identifier.authorityCheung, JPY=rp01685-
dc.identifier.authorityKwan, KYH=rp02014-
dc.identifier.authoritySze, KY=rp00171-
dc.identifier.authorityCheung, KMC=rp00387-
dc.description.naturepostprint-
dc.identifier.doi10.1097/BRS.0000000000003268-
dc.identifier.pmid31574056-
dc.identifier.scopuseid_2-s2.0-85080835741-
dc.identifier.hkuros308652-
dc.identifier.volume45-
dc.identifier.issue6-
dc.identifier.spageE346-
dc.identifier.epageE351-
dc.identifier.isiWOS:000544932100007-
dc.publisher.placeUnited States-
dc.identifier.issnl0362-2436-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats