File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: An HPLC-DAD method for rapid and high resolution analysis of concentrated BTEX and styrene aqueous samples

TitleAn HPLC-DAD method for rapid and high resolution analysis of concentrated BTEX and styrene aqueous samples
Authors
Issue Date2012
Citation
Analytical Methods, 2012, v. 4, n. 11, p. 3545-3550 How to Cite?
AbstractA rapid HPLC-DAD method for analysis of concentrated BTEX and styrene (BTEXS) aqueous mixtures is reported. Good resolutions of close to or greater than 1.5 were obtained for high equimolar BTEXS concentrations of up to 2.0 mM. At 5.5 min per sample analysis, this method is also one of the fastest HPLC methods to date, providing high throughput analysis and lowering the analysis price. © 2012 The Royal Society of Chemistry.
Persistent Identifierhttp://hdl.handle.net/10722/270335
ISSN
2021 Impact Factor: 3.532
2020 SCImago Journal Rankings: 0.615
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorAmy Tan, Giin Yu-
dc.contributor.authorChen, Chia Lung-
dc.contributor.authorZhao, Lei-
dc.contributor.authorMo, Yu-
dc.contributor.authorChang, Victor W.C.-
dc.contributor.authorWang, Jing Yuan-
dc.date.accessioned2019-05-27T03:57:19Z-
dc.date.available2019-05-27T03:57:19Z-
dc.date.issued2012-
dc.identifier.citationAnalytical Methods, 2012, v. 4, n. 11, p. 3545-3550-
dc.identifier.issn1759-9660-
dc.identifier.urihttp://hdl.handle.net/10722/270335-
dc.description.abstractA rapid HPLC-DAD method for analysis of concentrated BTEX and styrene (BTEXS) aqueous mixtures is reported. Good resolutions of close to or greater than 1.5 were obtained for high equimolar BTEXS concentrations of up to 2.0 mM. At 5.5 min per sample analysis, this method is also one of the fastest HPLC methods to date, providing high throughput analysis and lowering the analysis price. © 2012 The Royal Society of Chemistry.-
dc.languageeng-
dc.relation.ispartofAnalytical Methods-
dc.titleAn HPLC-DAD method for rapid and high resolution analysis of concentrated BTEX and styrene aqueous samples-
dc.typeArticle-
dc.description.naturelink_to_subscribed_fulltext-
dc.identifier.doi10.1039/c2ay25947h-
dc.identifier.scopuseid_2-s2.0-84868156536-
dc.identifier.volume4-
dc.identifier.issue11-
dc.identifier.spage3545-
dc.identifier.epage3550-
dc.identifier.eissn1759-9679-
dc.identifier.isiWOS:000310281700007-
dc.identifier.issnl1759-9660-

Export via OAI-PMH Interface in XML Formats