File Download
There are no files associated with this item.
Links for fulltext
(May Require Subscription)
- Publisher Website: 10.1109/19.85345
- Scopus: eid_2-s2.0-0026204853
- WOS: WOS:A1991GB24600014
- Find via
Supplementary
- Citations:
- Appears in Collections:
Article: Improved calibration of a large open-ended coaxial probe for dielectric measurements
Title | Improved calibration of a large open-ended coaxial probe for dielectric measurements |
---|---|
Authors | |
Issue Date | 1991 |
Citation | Ieee Transactions On Instrumentation And Measurement, 1991, v. 40 n. 4, p. 742-746 How to Cite? |
Abstract | An improved method for calibrating an open-ended coaxial sensor is proposed. The calibration utilizes the measurements of an open, a short, and a short-cavity. These terminations are inexpensive to make and result in a calibration that depends only on the geometry of the sensor. The short-cavity is modeled by waveguide modal techniques and the theoretical reflection coefficient determined to any accuracy. Consequently, an improvement in the accuracy of the measurements is obtained over a large bandwidth. The calibration was tested by comparing theoretical and experimental values of the complex permittivity for common liquids, and the discrepancy was seen to be small. |
Persistent Identifier | http://hdl.handle.net/10722/182515 |
ISSN | 2023 Impact Factor: 5.6 2023 SCImago Journal Rankings: 1.536 |
ISI Accession Number ID |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Otto, GP | en_US |
dc.contributor.author | Chew, WC | en_US |
dc.date.accessioned | 2013-05-02T05:15:41Z | - |
dc.date.available | 2013-05-02T05:15:41Z | - |
dc.date.issued | 1991 | en_US |
dc.identifier.citation | Ieee Transactions On Instrumentation And Measurement, 1991, v. 40 n. 4, p. 742-746 | en_US |
dc.identifier.issn | 0018-9456 | en_US |
dc.identifier.uri | http://hdl.handle.net/10722/182515 | - |
dc.description.abstract | An improved method for calibrating an open-ended coaxial sensor is proposed. The calibration utilizes the measurements of an open, a short, and a short-cavity. These terminations are inexpensive to make and result in a calibration that depends only on the geometry of the sensor. The short-cavity is modeled by waveguide modal techniques and the theoretical reflection coefficient determined to any accuracy. Consequently, an improvement in the accuracy of the measurements is obtained over a large bandwidth. The calibration was tested by comparing theoretical and experimental values of the complex permittivity for common liquids, and the discrepancy was seen to be small. | en_US |
dc.language | eng | en_US |
dc.relation.ispartof | IEEE Transactions on Instrumentation and Measurement | en_US |
dc.title | Improved calibration of a large open-ended coaxial probe for dielectric measurements | en_US |
dc.type | Article | en_US |
dc.identifier.email | Chew, WC: wcchew@hku.hk | en_US |
dc.identifier.authority | Chew, WC=rp00656 | en_US |
dc.description.nature | link_to_subscribed_fulltext | en_US |
dc.identifier.doi | 10.1109/19.85345 | en_US |
dc.identifier.scopus | eid_2-s2.0-0026204853 | en_US |
dc.identifier.volume | 40 | en_US |
dc.identifier.issue | 4 | en_US |
dc.identifier.spage | 742 | en_US |
dc.identifier.epage | 746 | en_US |
dc.identifier.isi | WOS:A1991GB24600014 | - |
dc.publisher.place | United States | en_US |
dc.identifier.scopusauthorid | Otto, GP=35446329800 | en_US |
dc.identifier.scopusauthorid | Chew, WC=36014436300 | en_US |
dc.identifier.issnl | 0018-9456 | - |