File Download

There are no files associated with this item.

  Links for fulltext
     (May Require Subscription)
Supplementary

Article: A splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical results

TitleA splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical results
Authors
Issue Date1990
PublisherWB Saunders Co. The Journal's web site is located at http://www.elsevier.com/locate/jhsa
Citation
Journal Of Hand Surgery, 1990, v. 15 n. 4, p. 645-651 How to Cite?
AbstractA splint for controlled active motion after flexor tendon repair is described. It incorporates a single core-coated elastic band passing around a palmar pulley and attached proximally to a spring wire. Its mechanical properties were tested against six other systems. The tension in various systems all rose near full extension. However, the palmar pulley, the spring wire, and the elastic band each could lower the tension significantly. When the bending moments at the interphalangeal joints were measured, all systems produced a peak during the latter part of extension. With the palmar pulley, spring wire, and elastic band, the rise was minimal and in fact, the bending moments diminished near full extension. Initial results in 28 flexor tendon repairs using this splint showed less flexion contracture when compared with 78 flexor tendon repairs using a standard rubber band anchored at the wrist.
Persistent Identifierhttp://hdl.handle.net/10722/178176
ISSN
2021 Impact Factor: 2.342
2020 SCImago Journal Rankings: 1.882
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorChow, SPen_US
dc.contributor.authorStephens, MMen_US
dc.contributor.authorNgai, WKen_US
dc.contributor.authorPun, WKen_US
dc.contributor.authorCrosby So, CYCen_US
dc.contributor.authorChu, Men_US
dc.date.accessioned2012-12-19T09:43:15Z-
dc.date.available2012-12-19T09:43:15Z-
dc.date.issued1990en_US
dc.identifier.citationJournal Of Hand Surgery, 1990, v. 15 n. 4, p. 645-651en_US
dc.identifier.issn0363-5023en_US
dc.identifier.urihttp://hdl.handle.net/10722/178176-
dc.description.abstractA splint for controlled active motion after flexor tendon repair is described. It incorporates a single core-coated elastic band passing around a palmar pulley and attached proximally to a spring wire. Its mechanical properties were tested against six other systems. The tension in various systems all rose near full extension. However, the palmar pulley, the spring wire, and the elastic band each could lower the tension significantly. When the bending moments at the interphalangeal joints were measured, all systems produced a peak during the latter part of extension. With the palmar pulley, spring wire, and elastic band, the rise was minimal and in fact, the bending moments diminished near full extension. Initial results in 28 flexor tendon repairs using this splint showed less flexion contracture when compared with 78 flexor tendon repairs using a standard rubber band anchored at the wrist.en_US
dc.languageengen_US
dc.publisherWB Saunders Co. The Journal's web site is located at http://www.elsevier.com/locate/jhsaen_US
dc.relation.ispartofJournal of Hand Surgeryen_US
dc.subject.meshBiomechanicsen_US
dc.subject.meshContracture - Prevention & Controlen_US
dc.subject.meshEquipment Designen_US
dc.subject.meshFingers - Physiopathology - Surgeryen_US
dc.subject.meshHand - Physiopathology - Surgeryen_US
dc.subject.meshHumansen_US
dc.subject.meshMovementen_US
dc.subject.meshPostoperative Perioden_US
dc.subject.meshSplintsen_US
dc.subject.meshTendons - Physiopathology - Surgeryen_US
dc.titleA splint for controlled active motion after flexor tendon repair. Design, mechanical testing, and preliminary clinical resultsen_US
dc.typeArticleen_US
dc.identifier.emailChow, SP: spchow@hku.hken_US
dc.identifier.authorityChow, SP=rp00064en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1016/S0363-5023(09)90030-2-
dc.identifier.pmid2380531-
dc.identifier.scopuseid_2-s2.0-0025296667en_US
dc.identifier.volume15en_US
dc.identifier.issue4en_US
dc.identifier.spage645en_US
dc.identifier.epage651en_US
dc.identifier.isiWOS:A1990DP02600020-
dc.publisher.placeUnited Statesen_US
dc.identifier.scopusauthoridChow, SP=7201828376en_US
dc.identifier.scopusauthoridStephens, MM=7201575245en_US
dc.identifier.scopusauthoridNgai, WK=6602875630en_US
dc.identifier.scopusauthoridPun, WK=7003726073en_US
dc.identifier.scopusauthoridCrosby So, CYC=6504103502en_US
dc.identifier.scopusauthoridChu, M=7402456181en_US
dc.identifier.issnl0363-5023-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats