Article: Comparison of the Williams Airway Intubator of Ovassapian Fibreoptic Intubating Airway for fibreoptic orotracheal intubation

TitleComparison of the Williams Airway Intubator of Ovassapian Fibreoptic Intubating Airway for fibreoptic orotracheal intubation
Authors
KeywordsEquipment; Ovassapian airway, Williams Airway Intubator, bronchoscope
Intubation, tracheal; fibreoptic
Issue Date2004
PublisherWiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044
Citation
Anaesthesia, 2004, v. 59 n. 2, p. 173-176 How to Cite?
AbstractA randomised study was carried out in 60 anaesthetised patients with no evidence of airway difficulties, to compare the Williams Airway Intubator and the Ovassapian Fibreoptic Intubating Airway for bronchoscopic view and ease of railroading a tracheal tube during fibreoptic orotracheal intubation. The Williams Airway Intubator provided a better view of the glottis (41 (68.3%) unobstructed views) than the Ovassapian Fibreoptic Intubating Airway (15 (25%) unobstructed views; p < 0.0001) and a shorter time was needed to complete bronchoscopy (median (95% CI) difference 4 s (1-7); p = 0.01). Four (6.7%) bronchoscopies failed using the Williams Airway Intubator compared with 26 (43.3%) using the Ovassapian Fibreoptic Intubating Airway (p < 0.0001). Both airways provided similar intubating conditions when the glottis was visible.
Persistent Identifierhttp://hdl.handle.net/10722/147207
ISSN
2021 Impact Factor: 12.893
2020 SCImago Journal Rankings: 1.839
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorGreenland, KBen_US
dc.contributor.authorLam, MCen_US
dc.contributor.authorIrwin, MGen_US
dc.date.accessioned2012-05-29T06:00:47Z-
dc.date.available2012-05-29T06:00:47Z-
dc.date.issued2004en_US
dc.identifier.citationAnaesthesia, 2004, v. 59 n. 2, p. 173-176en_US
dc.identifier.issn0003-2409en_US
dc.identifier.urihttp://hdl.handle.net/10722/147207-
dc.description.abstractA randomised study was carried out in 60 anaesthetised patients with no evidence of airway difficulties, to compare the Williams Airway Intubator and the Ovassapian Fibreoptic Intubating Airway for bronchoscopic view and ease of railroading a tracheal tube during fibreoptic orotracheal intubation. The Williams Airway Intubator provided a better view of the glottis (41 (68.3%) unobstructed views) than the Ovassapian Fibreoptic Intubating Airway (15 (25%) unobstructed views; p < 0.0001) and a shorter time was needed to complete bronchoscopy (median (95% CI) difference 4 s (1-7); p = 0.01). Four (6.7%) bronchoscopies failed using the Williams Airway Intubator compared with 26 (43.3%) using the Ovassapian Fibreoptic Intubating Airway (p < 0.0001). Both airways provided similar intubating conditions when the glottis was visible.en_US
dc.languageengen_US
dc.publisherWiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044en_US
dc.relation.ispartofAnaesthesiaen_US
dc.subjectEquipment; Ovassapian airway, Williams Airway Intubator, bronchoscope-
dc.subjectIntubation, tracheal; fibreoptic-
dc.subject.meshAdolescenten_US
dc.subject.meshAdulten_US
dc.subject.meshAgeden_US
dc.subject.meshAged, 80 And Overen_US
dc.subject.meshAnesthesia, Generalen_US
dc.subject.meshBronchoscopesen_US
dc.subject.meshFemaleen_US
dc.subject.meshFiber Optic Technology - Instrumentationen_US
dc.subject.meshGlottisen_US
dc.subject.meshHumansen_US
dc.subject.meshIntubation, Intratracheal - Instrumentationen_US
dc.subject.meshMaleen_US
dc.subject.meshMiddle Ageden_US
dc.subject.meshOdds Ratioen_US
dc.subject.meshTime Factorsen_US
dc.titleComparison of the Williams Airway Intubator of Ovassapian Fibreoptic Intubating Airway for fibreoptic orotracheal intubationen_US
dc.typeArticleen_US
dc.identifier.emailIrwin, MG:mgirwin@hku.hken_US
dc.identifier.authorityIrwin, MG=rp00390en_US
dc.description.naturelink_to_OA_fulltexten_US
dc.identifier.doi10.1111/j.1365-2044.2004.03527.xen_US
dc.identifier.pmid14725520-
dc.identifier.scopuseid_2-s2.0-0442325407en_US
dc.identifier.hkuros89746-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0442325407&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume59en_US
dc.identifier.issue2en_US
dc.identifier.spage173en_US
dc.identifier.epage176en_US
dc.identifier.isiWOS:000188226600011-
dc.publisher.placeUnited Kingdomen_US
dc.identifier.issnl0003-2409-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats