Conference Paper: Evaluation of a Degenerated Birdcage Coil for Parallel Imaging
Title | Evaluation of a Degenerated Birdcage Coil for Parallel Imaging |
---|---|
Authors | |
Issue Date | 2007 |
Publisher | International Society for Magnetic Resonance in Medicine |
Citation | Joint Annual Meeting ISMRM-ESMRMB SMRT16th Annual Meeting, Berlin, Germany, 19-25 May 2007, Abstract no. 1039 How to Cite? |
Abstract | The higher-order modes of a 12-rung degenerated low-pass birdcage coil for parallel imaging has been evaluated using FDTD simulations in a spherical
phantom at 3T. When all 6 order modes are incorporated, the acceleration factor can be up to 4 with average g-factor less than 1.8 and maximum g-factor
less than 3.6. For certain number of channels, the optimized combination of corresponding modes can be obtained. For a specific number of receive channel
in MRI system, the optimum combination of modes can be obtained based on different acceleration factors. |
Persistent Identifier | http://hdl.handle.net/10722/99762 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Wang, C | en_HK |
dc.contributor.author | Qu, P | en_HK |
dc.contributor.author | Yuan, J | en_HK |
dc.contributor.author | Shen, GG | en_HK |
dc.date.accessioned | 2010-09-25T18:43:12Z | - |
dc.date.available | 2010-09-25T18:43:12Z | - |
dc.date.issued | 2007 | en_HK |
dc.identifier.citation | Joint Annual Meeting ISMRM-ESMRMB SMRT16th Annual Meeting, Berlin, Germany, 19-25 May 2007, Abstract no. 1039 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/99762 | - |
dc.description.abstract | The higher-order modes of a 12-rung degenerated low-pass birdcage coil for parallel imaging has been evaluated using FDTD simulations in a spherical phantom at 3T. When all 6 order modes are incorporated, the acceleration factor can be up to 4 with average g-factor less than 1.8 and maximum g-factor less than 3.6. For certain number of channels, the optimized combination of corresponding modes can be obtained. For a specific number of receive channel in MRI system, the optimum combination of modes can be obtained based on different acceleration factors. | - |
dc.language | eng | en_HK |
dc.publisher | International Society for Magnetic Resonance in Medicine | - |
dc.relation.ispartof | Joint Annual Meeting ISMRM-ESMRMB SMRT | en_HK |
dc.title | Evaluation of a Degenerated Birdcage Coil for Parallel Imaging | en_HK |
dc.type | Conference_Paper | en_HK |
dc.identifier.email | Shen, GG: gxshen@eee.hku.hk | en_HK |
dc.identifier.authority | Shen, GG=rp00166 | en_HK |
dc.identifier.hkuros | 132660 | en_HK |
dc.identifier.spage | 1039 | en_HK |