File Download
Supplementary

Conference Paper: Comparison of Three Applied Volume Coils at 7T

TitleComparison of Three Applied Volume Coils at 7T
Authors
Issue Date2006
PublisherThe International Society for Magnetic Resonance in Medicine (ISMRM)
Citation
The International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 3533 How to Cite?
AbstractThe B1 + field homogeneity, SAR and SNR of shielded birdcage, TEM and microstrip volume coils are compared at 7T in this work by finite difference time domain (FDTD) method. It seems that the birdcage coil has the best unload B1 + homogeneity but the TEM has the best loaded B1 + homogeneity. On the other hand, the microstrip coil seems better in terms of SAR and average SNR than other two coils at ultra high field MRI.
Persistent Identifierhttp://hdl.handle.net/10722/99665

 

DC FieldValueLanguage
dc.contributor.authorWang, Cen_HK
dc.contributor.authorYuan, Jen_HK
dc.contributor.authorQu, Pen_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:39:28Z-
dc.date.available2010-09-25T18:39:28Z-
dc.date.issued2006en_HK
dc.identifier.citationThe International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 3533en_HK
dc.identifier.urihttp://hdl.handle.net/10722/99665-
dc.description.abstractThe B1 + field homogeneity, SAR and SNR of shielded birdcage, TEM and microstrip volume coils are compared at 7T in this work by finite difference time domain (FDTD) method. It seems that the birdcage coil has the best unload B1 + homogeneity but the TEM has the best loaded B1 + homogeneity. On the other hand, the microstrip coil seems better in terms of SAR and average SNR than other two coils at ultra high field MRI.-
dc.languageengen_HK
dc.publisherThe International Society for Magnetic Resonance in Medicine (ISMRM)-
dc.relation.ispartofThe International Society for Magnetic Resonance in Medicine Scientific Meeting & Exhibition (ISMRM)en_HK
dc.titleComparison of Three Applied Volume Coils at 7Ten_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.description.naturepublished_or_final_version-
dc.identifier.hkuros121318en_HK
dc.identifier.spage3533en_HK

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats