File Download
Supplementary
-
Citations:
- Appears in Collections:
Conference Paper: Modified microstrip volume coil for ultrahigh field
Title | Modified microstrip volume coil for ultrahigh field |
---|---|
Authors | |
Issue Date | 2005 |
Publisher | International Society for Magnetic Resonance in Medicine. |
Citation | International Society for Magnetic Resonance in Medicine 13th Scientific Meeting & Exhibition, Miami, FL, 7-13 May 2005, p. 947 How to Cite? |
Abstract | It is a challenge to design a lumped-element microstrip volume resonator at ultrahigh field (>7T) with large size. A new modified microstrip volume coil
with large coil dimension (o.d.=18cm) and coil thickness (3.3cm) is fabricated, tested and simulated in this paper. Through deploying ground line at the back
of each strip line, this modified microstrip resonator can be easily tuned to 300MHz. |
Persistent Identifier | http://hdl.handle.net/10722/99604 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Wu, B | en_HK |
dc.contributor.author | Wei, J | en_HK |
dc.contributor.author | Wang, C | en_HK |
dc.contributor.author | Pang, Y | en_HK |
dc.contributor.author | Shen, GG | en_HK |
dc.date.accessioned | 2010-09-25T18:37:05Z | - |
dc.date.available | 2010-09-25T18:37:05Z | - |
dc.date.issued | 2005 | en_HK |
dc.identifier.citation | International Society for Magnetic Resonance in Medicine 13th Scientific Meeting & Exhibition, Miami, FL, 7-13 May 2005, p. 947 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/99604 | - |
dc.description.abstract | It is a challenge to design a lumped-element microstrip volume resonator at ultrahigh field (>7T) with large size. A new modified microstrip volume coil with large coil dimension (o.d.=18cm) and coil thickness (3.3cm) is fabricated, tested and simulated in this paper. Through deploying ground line at the back of each strip line, this modified microstrip resonator can be easily tuned to 300MHz. | - |
dc.language | eng | en_HK |
dc.publisher | International Society for Magnetic Resonance in Medicine. | en_HK |
dc.relation.ispartof | International Society for Magnetic Resonance in Medicine Scientific Meeting & Exhibition | en_HK |
dc.title | Modified microstrip volume coil for ultrahigh field | en_HK |
dc.type | Conference_Paper | en_HK |
dc.identifier.email | Shen, GG: gxshen@eee.hku.hk | en_HK |
dc.identifier.authority | Shen, GG=rp00166 | en_HK |
dc.description.nature | published_or_final_version | - |
dc.identifier.hkuros | 104539 | en_HK |
dc.identifier.volume | , v. 1 | en_HK |
dc.identifier.spage | 947 | en_HK |