Conference Paper: Noise Figure and Dynamic Range Optimization in Optical Links for MRI Applications

TitleNoise Figure and Dynamic Range Optimization in Optical Links for MRI Applications
Authors
Issue Date2006
PublisherThe International Society for Magnetic Resonance in Medicine (ISMRM)
Citation
The International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 2031 How to Cite?
AbstractMRI applications impose critical requirements in noise figure and dynamic range for optical links used for coil interconnect. To reduce the intrinsic high noise figure of optical links, a low-noise preamp must be used. But it also brings the side effect of dynamic range reduction. Compromise must be made between noise figure and dynamic range to optimize the optical link to meet the critical requirement in MRI. Dynamic range of the current optical link is still a challenge for high field imaging above 3T and needs to be further improved.
Persistent Identifierhttp://hdl.handle.net/10722/99567

 

DC FieldValueLanguage
dc.contributor.authorYuan, Jen_HK
dc.contributor.authorQu, Pen_HK
dc.contributor.authorWei, Jen_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:35:39Z-
dc.date.available2010-09-25T18:35:39Z-
dc.date.issued2006en_HK
dc.identifier.citationThe International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 2031en_HK
dc.identifier.urihttp://hdl.handle.net/10722/99567-
dc.description.abstractMRI applications impose critical requirements in noise figure and dynamic range for optical links used for coil interconnect. To reduce the intrinsic high noise figure of optical links, a low-noise preamp must be used. But it also brings the side effect of dynamic range reduction. Compromise must be made between noise figure and dynamic range to optimize the optical link to meet the critical requirement in MRI. Dynamic range of the current optical link is still a challenge for high field imaging above 3T and needs to be further improved.-
dc.languageengen_HK
dc.publisherThe International Society for Magnetic Resonance in Medicine (ISMRM)-
dc.relation.ispartofThe International Society for Magnetic Resonance in Medicine Scientific Meeting & Exhibition (ISMRM)en_HK
dc.titleNoise Figure and Dynamic Range Optimization in Optical Links for MRI Applicationsen_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.description.naturepublished_or_final_version-
dc.identifier.hkuros121341en_HK
dc.identifier.spage2031en_HK