File Download
  Links for fulltext
     (May Require Subscription)
  • Find via Find It@HKUL
Supplementary

Conference Paper: Microstrip Loop Array (MLA) for Parallel Imaging

TitleMicrostrip Loop Array (MLA) for Parallel Imaging
Authors
Issue Date2004
PublisherThe International Society for Magnetic Resonance in Medicine (ISMRM)
Citation
The International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 1601 How to Cite?
AbstractThe microstrip loop array (MLA), an improved version of planar strip array, is proposed which take advantage of loop geometry. Without much loss of the well-decoupling benefit of strip line arrays, the proposed microstrip loop arrays have well-localized sensitivity profiles which are more suitable for parallel imaging. Simulation and experiment results show that microstrip loop coils have sensitivity profiles like conventional loop coils, while the mutual coupling is 10-20dB less at same distances. A sensitivity-encoded cardiac imaging was performed to demonstrate the suitability of MLA for parallel imaging.
Persistent Identifierhttp://hdl.handle.net/10722/99444
ISSN

 

DC FieldValueLanguage
dc.contributor.authorQu, Pen_HK
dc.contributor.authorWu, Ben_HK
dc.contributor.authorWei, Jen_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:30:29Z-
dc.date.available2010-09-25T18:30:29Z-
dc.date.issued2004en_HK
dc.identifier.citationThe International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 1601en_HK
dc.identifier.issn1545-4436-
dc.identifier.urihttp://hdl.handle.net/10722/99444-
dc.description.abstractThe microstrip loop array (MLA), an improved version of planar strip array, is proposed which take advantage of loop geometry. Without much loss of the well-decoupling benefit of strip line arrays, the proposed microstrip loop arrays have well-localized sensitivity profiles which are more suitable for parallel imaging. Simulation and experiment results show that microstrip loop coils have sensitivity profiles like conventional loop coils, while the mutual coupling is 10-20dB less at same distances. A sensitivity-encoded cardiac imaging was performed to demonstrate the suitability of MLA for parallel imaging.-
dc.languageengen_HK
dc.publisherThe International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.relation.ispartofProceedings of the International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.titleMicrostrip Loop Array (MLA) for Parallel Imagingen_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.description.naturepublished_or_final_version-
dc.identifier.hkuros91378en_HK
dc.identifier.volume11en_HK
dc.identifier.spage1601en_HK
dc.identifier.issnl1524-6965-

Export via OAI-PMH Interface in XML Formats