Conference Paper: A simple but robust isotropic and background gradient independent diffusion gradient design
Title | A simple but robust isotropic and background gradient independent diffusion gradient design |
---|---|
Authors | |
Issue Date | 2008 |
Publisher | International Society of Magnetic Resonance in Medicine |
Citation | International Society of Magnetic Resonance in Medicine (ISMRM) 16th Scientific Meeting and Exhibition, Toronto, Canada, 3-9 May 2008, p. 1816 How to Cite? |
Abstract | Due to computation and design complexity, there is no design so far for isotropic and background gradient independent diffusion gradients. In this abstract,
we present a simple but robust method to design orientation invariant and background magnetic field insensitive diffusion gradients. Although the diffusion
weighting efficiency is reduced, this design provides a set of diffusion gradients that may be used to map ADC efficiently and accurately or in presence of
severe field inhomogeneity such as induced by metal implants. |
Persistent Identifier | http://hdl.handle.net/10722/99411 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Guo, H | en_HK |
dc.contributor.author | Song, AW | en_HK |
dc.contributor.author | Wu, EX | en_HK |
dc.date.accessioned | 2010-09-25T18:28:59Z | - |
dc.date.available | 2010-09-25T18:28:59Z | - |
dc.date.issued | 2008 | en_HK |
dc.identifier.citation | International Society of Magnetic Resonance in Medicine (ISMRM) 16th Scientific Meeting and Exhibition, Toronto, Canada, 3-9 May 2008, p. 1816 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/99411 | - |
dc.description.abstract | Due to computation and design complexity, there is no design so far for isotropic and background gradient independent diffusion gradients. In this abstract, we present a simple but robust method to design orientation invariant and background magnetic field insensitive diffusion gradients. Although the diffusion weighting efficiency is reduced, this design provides a set of diffusion gradients that may be used to map ADC efficiently and accurately or in presence of severe field inhomogeneity such as induced by metal implants. | - |
dc.language | eng | en_HK |
dc.publisher | International Society of Magnetic Resonance in Medicine | - |
dc.relation.ispartof | 2008 Proceedings of International Society of Magnetic Resonance in Medicine | en_HK |
dc.title | A simple but robust isotropic and background gradient independent diffusion gradient design | en_HK |