Conference Paper: Inductively Decoupled Microstrip Array at 9.4T
Title | Inductively Decoupled Microstrip Array at 9.4T |
---|---|
Authors | |
Issue Date | 2006 |
Publisher | The International Society for Magnetic Resonance in Medicine (ISMRM) |
Citation | The International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 528 How to Cite? |
Abstract | In this study, an inductive decoupling technique of microstrip array for ultra high fields has been discussed. The decoupling inductance is nearly
independent of resonant frequency. Thus microstrip arrays can be tuned and work well at the magnetic fields higher than 7T. For the prototype of volume
array, the introduced inductors can reduce the mutual coupling not only between adjacent elements, but also between non-adjacent elements. |
Persistent Identifier | http://hdl.handle.net/10722/99214 |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Wu, B | en_HK |
dc.contributor.author | Zhang, X | en_HK |
dc.contributor.author | Shen, GG | en_HK |
dc.date.accessioned | 2010-09-25T18:20:30Z | - |
dc.date.available | 2010-09-25T18:20:30Z | - |
dc.date.issued | 2006 | en_HK |
dc.identifier.citation | The International Society for Magnetic Resonance in Medicine 14th Scientific Meeting & Exhibition (ISMRM2006). Seattle, WA, 6-12 May 2006, p. 528 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/99214 | - |
dc.description.abstract | In this study, an inductive decoupling technique of microstrip array for ultra high fields has been discussed. The decoupling inductance is nearly independent of resonant frequency. Thus microstrip arrays can be tuned and work well at the magnetic fields higher than 7T. For the prototype of volume array, the introduced inductors can reduce the mutual coupling not only between adjacent elements, but also between non-adjacent elements. | - |
dc.language | eng | en_HK |
dc.publisher | The International Society for Magnetic Resonance in Medicine (ISMRM) | - |
dc.relation.ispartof | The International Society for Magnetic Resonance in Medicine Scientific Meeting & Exhibition (ISMRM) | en_HK |
dc.title | Inductively Decoupled Microstrip Array at 9.4T | en_HK |
dc.type | Conference_Paper | en_HK |
dc.identifier.email | Shen, GG: gxshen@eee.hku.hk | en_HK |
dc.identifier.authority | Shen, GG=rp00166 | en_HK |
dc.description.nature | published_or_final_version | - |
dc.identifier.hkuros | 121290 | en_HK |
dc.identifier.spage | 528 | en_HK |