Conference Paper: A shielding-based decoupling technique for coil array design

TitleA shielding-based decoupling technique for coil array design
Authors
Issue Date2004
PublisherThe International Society for Magnetic Resonance in Medicine (ISMRM)
Citation
The International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 1605 How to Cite?
AbstractA new decoupling technique which takes advantage of shielding for coil array design is proposed. A ground plane is used to reduce mutual inductance by reducing the effective coupling loop area of circuits; in addition, a "screen track" is placed between adjacent coils to minimize magnetic interferences. This technique requires no modification of circuit and doesnít cause loss of flexibility of coil geometry. Simulation and experiments show that a ground plane beneath surface coils can reduce mutual coupling by 10-20dB, and a "screen track" can further provide a 2-8dB more isolation.
Persistent Identifierhttp://hdl.handle.net/10722/99191
ISSN

 

DC FieldValueLanguage
dc.contributor.authorQu, Pen_HK
dc.contributor.authorWu, Ben_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:19:33Z-
dc.date.available2010-09-25T18:19:33Z-
dc.date.issued2004en_HK
dc.identifier.citationThe International Society for Magnetic Resonance in Medicine (ISMRM) 12th Scientific Meeting & Exhibition, Kyoto, Japan, 15-21 May 2004. In Proceedings of the International Society for Magnetic Resonance in Medicine (ISMRM), 2004, v. 11, p. 1605en_HK
dc.identifier.issn1545-4436-
dc.identifier.urihttp://hdl.handle.net/10722/99191-
dc.description.abstractA new decoupling technique which takes advantage of shielding for coil array design is proposed. A ground plane is used to reduce mutual inductance by reducing the effective coupling loop area of circuits; in addition, a "screen track" is placed between adjacent coils to minimize magnetic interferences. This technique requires no modification of circuit and doesnít cause loss of flexibility of coil geometry. Simulation and experiments show that a ground plane beneath surface coils can reduce mutual coupling by 10-20dB, and a "screen track" can further provide a 2-8dB more isolation.-
dc.languageengen_HK
dc.publisherThe International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.relation.ispartofProceedings of the International Society for Magnetic Resonance in Medicine (ISMRM)en_HK
dc.titleA shielding-based decoupling technique for coil array designen_HK
dc.typeConference_Paperen_HK
dc.identifier.emailShen, GG: gxshen@eee.hku.hken_HK
dc.identifier.authorityShen, GG=rp00166en_HK
dc.description.naturepublished_or_final_version-
dc.identifier.hkuros91380en_HK
dc.identifier.volume11en_HK
dc.identifier.spage1605en_HK
dc.identifier.issnl1524-6965-