Conference Paper: Investigation of dynamic range requirement for MRI signal transmission by optical fiber link

TitleInvestigation of dynamic range requirement for MRI signal transmission by optical fiber link
Authors
Issue Date2007
PublisherInternational Society for Magnetic Resonance in Medicine
Citation
Joint Annual Meeting ISMRM-ESMRMB SMRT16th Annual Meeting, Berlin, Germany, 19-25 May 2007, Abstract no. 995 How to Cite?
AbstractLink dynamic range is an important factor for MRI signal transmission by optical fiber links. Dynamic range of a direct modulation optical fiber link is analyzed theoretically and tested on both 0.3T and 1.5T MRI. It is found that the optical link has wide enough dynamic range to transmit MRI signal at 0.3T. However, much wider dynamic range is required at 1.5T. This requirement can be compensated by the use of low noise preamp and attenuators to avoid suppression and distortion.
Persistent Identifierhttp://hdl.handle.net/10722/99146

 

DC FieldValueLanguage
dc.contributor.authorYuan, Jen_HK
dc.contributor.authorWei, Jen_HK
dc.contributor.authorDu, C.en_HK
dc.contributor.authorShen, GGen_HK
dc.date.accessioned2010-09-25T18:17:43Z-
dc.date.available2010-09-25T18:17:43Z-
dc.date.issued2007en_HK
dc.identifier.citationJoint Annual Meeting ISMRM-ESMRMB SMRT16th Annual Meeting, Berlin, Germany, 19-25 May 2007, Abstract no. 995en_HK
dc.identifier.urihttp://hdl.handle.net/10722/99146-
dc.description.abstractLink dynamic range is an important factor for MRI signal transmission by optical fiber links. Dynamic range of a direct modulation optical fiber link is analyzed theoretically and tested on both 0.3T and 1.5T MRI. It is found that the optical link has wide enough dynamic range to transmit MRI signal at 0.3T. However, much wider dynamic range is required at 1.5T. This requirement can be compensated by the use of low noise preamp and attenuators to avoid suppression and distortion.-
dc.languageengen_HK
dc.publisherInternational Society for Magnetic Resonance in Medicine-
dc.relation.ispartofJoint Annual Meeting of ISMRM-ESMRMB SMRTen_HK
dc.titleInvestigation of dynamic range requirement for MRI signal transmission by optical fiber linken_HK