File Download
  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Cutting errors in total knee replacement: Assessment by computer assisted surgery

TitleCutting errors in total knee replacement: Assessment by computer assisted surgery
Authors
KeywordsCutting error
Total knee replacement
Issue Date2008
PublisherSpringer. The Journal's web site is located at http://www.springer.com/medicine/orthopedics/journal/167
Citation
Knee Surgery, Sports Traumatology, Arthroscopy, 2008, v. 16 n. 7, p. 670-673 How to Cite?
AbstractThe observed errors in the position of the implanted prosthesis can be due to a number of potential causes. One of these is the potential error during execution of the bone cuts. However, there is only minimal information on this in the current literature. The amount of cutting errors in 40 consecutive total knee replacements was reported. All the operations were done by the same surgeon. The amount of cutting error was measured by the use of computer navigation system. It was hypothesized that there was no difference in the amount of error between bone cut through the cutting slot (slotted cutting) and bone cut done on the surface of the cutting guide (open cutting). It was found that the average absolute cutting error was 1° in the coronal plane and 1.4° in the sagittal plane. Significantly more outlier (more than 3°) was observed in the errors in the sagittal plane (P = 0.014, chi square test). Open cutting resulted in less error in the sagittal plane of the tibial cut when compared with slotted cutting (P = 0.031, Mann-Whitney U Test). This was attributed by the use of a thicker saw blade with higher stiffness in the open cutting method. © 2008 Springer-Verlag.
Persistent Identifierhttp://hdl.handle.net/10722/79662
ISSN
2023 Impact Factor: 3.3
2023 SCImago Journal Rankings: 1.785
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorYau, WPen_HK
dc.contributor.authorChiu, KYen_HK
dc.date.accessioned2010-09-06T07:57:10Z-
dc.date.available2010-09-06T07:57:10Z-
dc.date.issued2008en_HK
dc.identifier.citationKnee Surgery, Sports Traumatology, Arthroscopy, 2008, v. 16 n. 7, p. 670-673en_HK
dc.identifier.issn0942-2056en_HK
dc.identifier.urihttp://hdl.handle.net/10722/79662-
dc.description.abstractThe observed errors in the position of the implanted prosthesis can be due to a number of potential causes. One of these is the potential error during execution of the bone cuts. However, there is only minimal information on this in the current literature. The amount of cutting errors in 40 consecutive total knee replacements was reported. All the operations were done by the same surgeon. The amount of cutting error was measured by the use of computer navigation system. It was hypothesized that there was no difference in the amount of error between bone cut through the cutting slot (slotted cutting) and bone cut done on the surface of the cutting guide (open cutting). It was found that the average absolute cutting error was 1° in the coronal plane and 1.4° in the sagittal plane. Significantly more outlier (more than 3°) was observed in the errors in the sagittal plane (P = 0.014, chi square test). Open cutting resulted in less error in the sagittal plane of the tibial cut when compared with slotted cutting (P = 0.031, Mann-Whitney U Test). This was attributed by the use of a thicker saw blade with higher stiffness in the open cutting method. © 2008 Springer-Verlag.en_HK
dc.languageengen_HK
dc.publisherSpringer. The Journal's web site is located at http://www.springer.com/medicine/orthopedics/journal/167en_HK
dc.relation.ispartofKnee Surgery, Sports Traumatology, Arthroscopyen_HK
dc.rightsThe original publication is available at www.springerlink.com-
dc.subjectCutting erroren_HK
dc.subjectTotal knee replacementen_HK
dc.subject.meshArthritis, Rheumatoid - surgery-
dc.subject.meshArthroplasty, Replacement, Knee - adverse effects-
dc.subject.meshMedical Errors - statistics and numerical data-
dc.subject.meshOsteoarthritis, Knee - surgery-
dc.subject.meshOsteotomy - adverse effects-
dc.titleCutting errors in total knee replacement: Assessment by computer assisted surgeryen_HK
dc.typeArticleen_HK
dc.identifier.openurlhttp://library.hku.hk:4550/resserv?sid=HKU:IR&issn=0942-2056&volume=16&issue=7&spage=670&epage=673&date=2008&atitle=Cutting+errors+in+total+knee+replacement:+assessment+by+computer+assisted+surgery.en_HK
dc.identifier.emailYau, WP:peterwpy@hkucc.hku.hken_HK
dc.identifier.emailChiu, KY:pkychiu@hkucc.hku.hken_HK
dc.identifier.authorityYau, WP=rp00500en_HK
dc.identifier.authorityChiu, KY=rp00379en_HK
dc.description.naturepostprint-
dc.identifier.doi10.1007/s00167-008-0550-xen_HK
dc.identifier.pmid18478202-
dc.identifier.scopuseid_2-s2.0-46149118422en_HK
dc.identifier.hkuros142876en_HK
dc.identifier.hkuros146358-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-46149118422&selection=ref&src=s&origin=recordpageen_HK
dc.identifier.volume16en_HK
dc.identifier.issue7en_HK
dc.identifier.spage670en_HK
dc.identifier.epage673en_HK
dc.identifier.isiWOS:000257227800009-
dc.publisher.placeGermanyen_HK
dc.identifier.scopusauthoridYau, WP=7005822441en_HK
dc.identifier.scopusauthoridChiu, KY=7202988127en_HK
dc.identifier.citeulike3095343-
dc.identifier.issnl0942-2056-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats