Article: A sheath-flow nanoelectrospray interface of microchip electrophoresis MS for glycoprotein and glycopeptide analysis
Title | A sheath-flow nanoelectrospray interface of microchip electrophoresis MS for glycoprotein and glycopeptide analysis |
---|---|
Authors | |
Keywords | Glycopeptides Glycoprotein Microchip electrophoresis MS Nanoelectrospray interface |
Issue Date | 2006 |
Publisher | Wiley - V C H Verlag GmbH & Co KGaA. |
Citation | Electrophoresis, 2006, v. 27 n. 24, p. 5059-5067 How to Cite? |
Abstract | Microchip was coupled with MS through a stable, sensitive, and controllable sheath-flow nanoelectrospray (nES) interface for glycoprotein and glycopeptide analysis. The nano-ESI (nESI) was made with a delivery capillary, a commercial nES capillary, and a stainless steel (SS) tube which were connected together through a tee unit. High voltage for nES was applied on the SS tube and the commercial nES capillary was used as nES emitter. The delivery capillary was attached to the microchannel for delivering liquid from microchip to the nESI source. The flow rate of sheath liquid was optimized to be 100-200 nL/ min which largely reduced the sample dilution. The detection limit of peptides on this microchip/MS platform was at femtomole level. Glycoprotein and glycopeptides were also successfully analyzed on the platform. All the glycoforms and glycopeptides of ribonuclease B (RNase B) were identified with this method. Some structures of the glycopeptides from RNase B were further characterized with MS/MS on the microchip, coupled with a quadrupole IT-MS. © 2006 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim. |
Persistent Identifier | http://hdl.handle.net/10722/69919 |
ISSN | 2023 Impact Factor: 3.0 2023 SCImago Journal Rankings: 0.541 |
ISI Accession Number ID | |
References |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Mao, X | en_HK |
dc.contributor.author | Chu, IK | en_HK |
dc.contributor.author | Lin, B | en_HK |
dc.date.accessioned | 2010-09-06T06:18:02Z | - |
dc.date.available | 2010-09-06T06:18:02Z | - |
dc.date.issued | 2006 | en_HK |
dc.identifier.citation | Electrophoresis, 2006, v. 27 n. 24, p. 5059-5067 | en_HK |
dc.identifier.issn | 0173-0835 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/69919 | - |
dc.description.abstract | Microchip was coupled with MS through a stable, sensitive, and controllable sheath-flow nanoelectrospray (nES) interface for glycoprotein and glycopeptide analysis. The nano-ESI (nESI) was made with a delivery capillary, a commercial nES capillary, and a stainless steel (SS) tube which were connected together through a tee unit. High voltage for nES was applied on the SS tube and the commercial nES capillary was used as nES emitter. The delivery capillary was attached to the microchannel for delivering liquid from microchip to the nESI source. The flow rate of sheath liquid was optimized to be 100-200 nL/ min which largely reduced the sample dilution. The detection limit of peptides on this microchip/MS platform was at femtomole level. Glycoprotein and glycopeptides were also successfully analyzed on the platform. All the glycoforms and glycopeptides of ribonuclease B (RNase B) were identified with this method. Some structures of the glycopeptides from RNase B were further characterized with MS/MS on the microchip, coupled with a quadrupole IT-MS. © 2006 Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim. | en_HK |
dc.language | eng | en_HK |
dc.publisher | Wiley - V C H Verlag GmbH & Co KGaA. | en_HK |
dc.relation.ispartof | Electrophoresis | en_HK |
dc.subject | Glycopeptides | en_HK |
dc.subject | Glycoprotein | en_HK |
dc.subject | Microchip electrophoresis | en_HK |
dc.subject | MS | en_HK |
dc.subject | Nanoelectrospray interface | en_HK |
dc.title | A sheath-flow nanoelectrospray interface of microchip electrophoresis MS for glycoprotein and glycopeptide analysis | en_HK |
dc.type | Article | en_HK |
dc.identifier.openurl | http://library.hku.hk:4550/resserv?sid=HKU:IR&issn=0173-0835&volume=27&spage=5059&epage=5067&date=2006&atitle=A+sheath-flow+nanoelectrospray+interface+of+microchip+electrophoresis+MS+for+glycoprotein+and+glycopeptide+analysis | en_HK |
dc.identifier.email | Chu, IK:ivankchu@hku.hk | en_HK |
dc.identifier.authority | Chu, IK=rp00683 | en_HK |
dc.description.nature | link_to_subscribed_fulltext | - |
dc.identifier.doi | 10.1002/elps.200600349 | en_HK |
dc.identifier.pmid | 17117389 | en_HK |
dc.identifier.scopus | eid_2-s2.0-33846031560 | en_HK |
dc.identifier.hkuros | 125257 | en_HK |
dc.relation.references | http://www.scopus.com/mlt/select.url?eid=2-s2.0-33846031560&selection=ref&src=s&origin=recordpage | en_HK |
dc.identifier.volume | 27 | en_HK |
dc.identifier.issue | 24 | en_HK |
dc.identifier.spage | 5059 | en_HK |
dc.identifier.epage | 5067 | en_HK |
dc.identifier.isi | WOS:000243304100022 | - |
dc.publisher.place | Germany | en_HK |
dc.identifier.scopusauthorid | Mao, X=7402841292 | en_HK |
dc.identifier.scopusauthorid | Chu, IK=7103327484 | en_HK |
dc.identifier.scopusauthorid | Lin, B=34668182800 | en_HK |
dc.identifier.issnl | 0173-0835 | - |