File Download
Supplementary
-
Citations:
- Appears in Collections:
Conference Paper: A real-time monitor using wavelet analysis of the Doppler heart sound for the detection of venous air embolism
Title | A real-time monitor using wavelet analysis of the Doppler heart sound for the detection of venous air embolism |
---|---|
Authors | |
Keywords | Electronics |
Issue Date | 1997 |
Publisher | IEEE. |
Citation | IEEE International Symposium on Circuits and Systems Proceedings, Hong Kong, China, 9-12 June 1997, v. 1, p. 189-192 How to Cite? |
Abstract | A fast detection algorithm for venous air embolism (VAE) was developed and implemented as a real-time monitor for detecting embolic heart sound and estimating embolic air volume. Its performance was evaluated under bolus injection of sub-clinical (0.0l to 0.80 ml) and continuous infusion of clinically significant (0.80 to 9.60 ml) air volumes in anaesthetized dogs. The clinically significant air emboli could be estimated based on the calibration curve obtained during sub-clinical VAE for a subject. The monitor also kept track of the cumulative embolic air volumes and alerted the anaesthetists once a predefined clinically significant embolic air volume was reached. As both humans and dogs share similar physiological conditions, our monitor for dogs are expected to be applicable to humans. |
Persistent Identifier | http://hdl.handle.net/10722/46011 |
ISSN |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Chan, BCB | en_HK |
dc.contributor.author | Chan, FHY | en_HK |
dc.contributor.author | Lui, PW | en_HK |
dc.contributor.author | Poon, PWF | en_HK |
dc.contributor.author | Lam, FK | en_HK |
dc.contributor.author | Wang, H | en_HK |
dc.date.accessioned | 2007-10-30T06:40:35Z | - |
dc.date.available | 2007-10-30T06:40:35Z | - |
dc.date.issued | 1997 | en_HK |
dc.identifier.citation | IEEE International Symposium on Circuits and Systems Proceedings, Hong Kong, China, 9-12 June 1997, v. 1, p. 189-192 | en_HK |
dc.identifier.issn | 0271-4302 | en_HK |
dc.identifier.uri | http://hdl.handle.net/10722/46011 | - |
dc.description.abstract | A fast detection algorithm for venous air embolism (VAE) was developed and implemented as a real-time monitor for detecting embolic heart sound and estimating embolic air volume. Its performance was evaluated under bolus injection of sub-clinical (0.0l to 0.80 ml) and continuous infusion of clinically significant (0.80 to 9.60 ml) air volumes in anaesthetized dogs. The clinically significant air emboli could be estimated based on the calibration curve obtained during sub-clinical VAE for a subject. The monitor also kept track of the cumulative embolic air volumes and alerted the anaesthetists once a predefined clinically significant embolic air volume was reached. As both humans and dogs share similar physiological conditions, our monitor for dogs are expected to be applicable to humans. | en_HK |
dc.format.extent | 459073 bytes | - |
dc.format.extent | 13817 bytes | - |
dc.format.extent | 8841 bytes | - |
dc.format.mimetype | application/pdf | - |
dc.format.mimetype | text/plain | - |
dc.format.mimetype | text/plain | - |
dc.language | eng | en_HK |
dc.publisher | IEEE. | en_HK |
dc.rights | ©1997 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE. | - |
dc.subject | Electronics | en_HK |
dc.title | A real-time monitor using wavelet analysis of the Doppler heart sound for the detection of venous air embolism | en_HK |
dc.type | Conference_Paper | en_HK |
dc.identifier.openurl | http://library.hku.hk:4550/resserv?sid=HKU:IR&issn=0271-4302&volume=1&spage=189&epage=192&date=1997&atitle=A+real-time+monitor+using+wavelet+analysis+of+the+Doppler+heart+sound+for+the+detection+of+venous+air+embolism | en_HK |
dc.description.nature | published_or_final_version | en_HK |
dc.identifier.doi | 10.1109/ISCAS.1997.608662 | en_HK |
dc.identifier.hkuros | 27190 | - |
dc.identifier.issnl | 0271-4302 | - |