Article: Fixation for comminuted phalangeal fractures: A biomechanical study of five methods

TitleFixation for comminuted phalangeal fractures: A biomechanical study of five methods
Authors
Issue Date1996
PublisherSage Publications Ltd.. The Journal's web site is located at http://jhs.sagepub.com
Citation
Journal Of Hand Surgery: European Volume, 1996, v. 21 n. 6, p. 765-767 How to Cite?
AbstractThe rigidities of five fixation methods have been studied with a comminuted phalangeal fracture model. Mechanical testing of compression, bending and torsion were performed for each fixation. Lateral plating with six screws seems to provide the most rigid fixation. If such lateral plating is not practicable, the four-Kirschner wire method would be a satisfactory alternative. © 1996 The British Society for Surgery of the Hand.
Persistent Identifierhttp://hdl.handle.net/10722/170005
ISSN
2021 Impact Factor: 2.206
2020 SCImago Journal Rankings: 1.104
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorLU, WWen_HK
dc.contributor.authorFurumachI, Ken_HK
dc.contributor.authorIP, WYen_HK
dc.contributor.authorChow, SPen_HK
dc.date.accessioned2012-10-30T06:04:40Z-
dc.date.available2012-10-30T06:04:40Z-
dc.date.issued1996en_HK
dc.identifier.citationJournal Of Hand Surgery: European Volume, 1996, v. 21 n. 6, p. 765-767en_HK
dc.identifier.issn1753-1934en_HK
dc.identifier.urihttp://hdl.handle.net/10722/170005-
dc.description.abstractThe rigidities of five fixation methods have been studied with a comminuted phalangeal fracture model. Mechanical testing of compression, bending and torsion were performed for each fixation. Lateral plating with six screws seems to provide the most rigid fixation. If such lateral plating is not practicable, the four-Kirschner wire method would be a satisfactory alternative. © 1996 The British Society for Surgery of the Hand.en_HK
dc.languageengen_US
dc.publisherSage Publications Ltd.. The Journal's web site is located at http://jhs.sagepub.comen_HK
dc.relation.ispartofJournal of Hand Surgery: European Volumeen_HK
dc.titleFixation for comminuted phalangeal fractures: A biomechanical study of five methodsen_HK
dc.typeArticleen_HK
dc.identifier.emailLU, WW: wwlu@hku.hken_HK
dc.identifier.emailIP, WY: wyip@hku.hken_HK
dc.identifier.emailChow, SP: spchow@hku.hken_HK
dc.identifier.authorityLU, WW=rp00411en_HK
dc.identifier.authorityIP, WY=rp00401en_HK
dc.identifier.authorityChow, SP=rp00064en_HK
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1016/S0266-7681(96)80183-5en_HK
dc.identifier.pmid8982920-
dc.identifier.scopuseid_2-s2.0-0029729747en_HK
dc.identifier.hkuros25226-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0029729747&selection=ref&src=s&origin=recordpageen_HK
dc.identifier.volume21en_HK
dc.identifier.issue6en_HK
dc.identifier.spage765en_HK
dc.identifier.epage767en_HK
dc.identifier.isiWOS:A1996VZ71300010-
dc.publisher.placeUnited Kingdomen_HK
dc.identifier.scopusauthoridLU, WW=7404215221en_HK
dc.identifier.scopusauthoridFurumachI, K=6506036172en_HK
dc.identifier.scopusauthoridIP, WY=35549641700en_HK
dc.identifier.scopusauthoridChow, SP=7201828376en_HK
dc.identifier.issnl1753-1934-