Article: Shortening of the sensed AV delay of a dual chamber pacemaker during normal sinus rhythm

TitleShortening of the sensed AV delay of a dual chamber pacemaker during normal sinus rhythm
Authors
KeywordsAtrial tachyarrhythmia
Automatic mode switching
Dual chamber pacemaker
Pacemaker atrioventricular delay
Issue Date2003
PublisherSpringer New York LLC. The Journal's web site is located at http://springerlink.metapress.com/openurl.asp?genre=journal&issn=1383-875X
Citation
Journal Of Interventional Cardiac Electrophysiology, 2003, v. 9 n. 1, p. 29-34 How to Cite?
AbstractWe studied the Medtronic Thera and Kappa 400 dual chamber pacemakers to determine the causes and mechanism of shortening of the sensed AV delay during normal sinus rhythm. When the sensed AV delay shortens to less than its programmed value, it lengthens gradually and returns to its programmed duration after a number of pacing cycles. This behavior is linked to the relatively slow automatic mode switching algorithm (AMS) and was observed when a transient increase in the sensed atrial rate failed to reach the point where AMS was activated and also upon exit from the AMS mode.
Persistent Identifierhttp://hdl.handle.net/10722/162716
ISSN
2023 Impact Factor: 2.1
2023 SCImago Journal Rankings: 0.767
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorBarold, SSen_US
dc.contributor.authorLau, CPen_US
dc.contributor.authorTse, HFen_US
dc.contributor.authorHess, Men_US
dc.date.accessioned2012-09-05T05:22:45Z-
dc.date.available2012-09-05T05:22:45Z-
dc.date.issued2003en_US
dc.identifier.citationJournal Of Interventional Cardiac Electrophysiology, 2003, v. 9 n. 1, p. 29-34en_US
dc.identifier.issn1383-875Xen_US
dc.identifier.urihttp://hdl.handle.net/10722/162716-
dc.description.abstractWe studied the Medtronic Thera and Kappa 400 dual chamber pacemakers to determine the causes and mechanism of shortening of the sensed AV delay during normal sinus rhythm. When the sensed AV delay shortens to less than its programmed value, it lengthens gradually and returns to its programmed duration after a number of pacing cycles. This behavior is linked to the relatively slow automatic mode switching algorithm (AMS) and was observed when a transient increase in the sensed atrial rate failed to reach the point where AMS was activated and also upon exit from the AMS mode.en_US
dc.languageengen_US
dc.publisherSpringer New York LLC. The Journal's web site is located at http://springerlink.metapress.com/openurl.asp?genre=journal&issn=1383-875Xen_US
dc.relation.ispartofJournal of Interventional Cardiac Electrophysiologyen_US
dc.subjectAtrial tachyarrhythmia-
dc.subjectAutomatic mode switching-
dc.subjectDual chamber pacemaker-
dc.subjectPacemaker atrioventricular delay-
dc.subject.meshCardiac Pacing, Artificialen_US
dc.subject.meshElectrocardiographyen_US
dc.subject.meshHumansen_US
dc.subject.meshPacemaker, Artificialen_US
dc.titleShortening of the sensed AV delay of a dual chamber pacemaker during normal sinus rhythmen_US
dc.typeArticleen_US
dc.identifier.emailTse, HF:hftse@hkucc.hku.hken_US
dc.identifier.authorityTse, HF=rp00428en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1023/A:1025316419920en_US
dc.identifier.pmid12975568-
dc.identifier.scopuseid_2-s2.0-0141427582en_US
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0141427582&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume9en_US
dc.identifier.issue1en_US
dc.identifier.spage29en_US
dc.identifier.epage34en_US
dc.identifier.isiWOS:000184844400005-
dc.publisher.placeUnited Statesen_US
dc.identifier.scopusauthoridBarold, SS=7101800584en_US
dc.identifier.scopusauthoridLau, CP=7401968501en_US
dc.identifier.scopusauthoridTse, HF=7006070805en_US
dc.identifier.scopusauthoridHess, M=7201426543en_US
dc.identifier.issnl1383-875X-

Export via OAI-PMH Interface in XML Formats


OR