Article: Effect of electrode polarity on the energy required for transthoraic atrial defibrillation

TitleEffect of electrode polarity on the energy required for transthoraic atrial defibrillation
Authors
Issue Date1999
PublisherExcerpta Medica, Inc.. The Journal's web site is located at http://www.ajconline.org/
Citation
American Journal Of Cardiology, 1999, v. 84 n. 2, p. 228-230 How to Cite?
AbstractResults of this study demonstrate that anodal and cathodal configurations are associated with the same overall efficacy when using shocks up to 360 J in strength, but that the cathodal configuration is more likely to restore sinus rhythm using lower energy shocks. These data validate the conventional practice of using a cathodal configuration for atrial defibrillation.
Persistent Identifierhttp://hdl.handle.net/10722/162300
ISSN
2023 Impact Factor: 2.3
2023 SCImago Journal Rankings: 0.950
ISI Accession Number ID
References

 

DC FieldValueLanguage
dc.contributor.authorOral, Hen_US
dc.contributor.authorBrinkman, Ken_US
dc.contributor.authorPelosi, Fen_US
dc.contributor.authorFlemming, Men_US
dc.contributor.authorTse, HFen_US
dc.contributor.authorKim, MHen_US
dc.contributor.authorMichaud, GFen_US
dc.contributor.authorKnight, BPen_US
dc.contributor.authorGoyal, Ren_US
dc.contributor.authorStrickberger, SAen_US
dc.contributor.authorMorady, Fen_US
dc.date.accessioned2012-09-05T05:18:45Z-
dc.date.available2012-09-05T05:18:45Z-
dc.date.issued1999en_US
dc.identifier.citationAmerican Journal Of Cardiology, 1999, v. 84 n. 2, p. 228-230en_US
dc.identifier.issn0002-9149en_US
dc.identifier.urihttp://hdl.handle.net/10722/162300-
dc.description.abstractResults of this study demonstrate that anodal and cathodal configurations are associated with the same overall efficacy when using shocks up to 360 J in strength, but that the cathodal configuration is more likely to restore sinus rhythm using lower energy shocks. These data validate the conventional practice of using a cathodal configuration for atrial defibrillation.en_US
dc.languageengen_US
dc.publisherExcerpta Medica, Inc.. The Journal's web site is located at http://www.ajconline.org/en_US
dc.relation.ispartofAmerican Journal of Cardiologyen_US
dc.subject.meshAdolescenten_US
dc.subject.meshAdulten_US
dc.subject.meshAgeden_US
dc.subject.meshAged, 80 And Overen_US
dc.subject.meshAtrial Fibrillation - Therapyen_US
dc.subject.meshBody Weighten_US
dc.subject.meshClinical Trials As Topicen_US
dc.subject.meshElectric Countershock - Methodsen_US
dc.subject.meshElectricityen_US
dc.subject.meshElectrodesen_US
dc.subject.meshFemaleen_US
dc.subject.meshHumansen_US
dc.subject.meshMaleen_US
dc.subject.meshMiddle Ageden_US
dc.subject.meshRegression Analysisen_US
dc.titleEffect of electrode polarity on the energy required for transthoraic atrial defibrillationen_US
dc.typeArticleen_US
dc.identifier.emailTse, HF:hftse@hkucc.hku.hken_US
dc.identifier.authorityTse, HF=rp00428en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1016/S0002-9149(99)00241-6en_US
dc.identifier.pmid10426347-
dc.identifier.scopuseid_2-s2.0-0032803914en_US
dc.identifier.hkuros41431-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-0032803914&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume84en_US
dc.identifier.issue2en_US
dc.identifier.spage228en_US
dc.identifier.epage230en_US
dc.identifier.isiWOS:000081351600023-
dc.publisher.placeUnited Statesen_US
dc.identifier.scopusauthoridOral, H=7006855160en_US
dc.identifier.scopusauthoridBrinkman, K=7006148844en_US
dc.identifier.scopusauthoridPelosi, F=7004832991en_US
dc.identifier.scopusauthoridFlemming, M=7003616446en_US
dc.identifier.scopusauthoridTse, HF=7006070805en_US
dc.identifier.scopusauthoridKim, MH=7406090815en_US
dc.identifier.scopusauthoridMichaud, GF=7006729860en_US
dc.identifier.scopusauthoridKnight, BP=7201940628en_US
dc.identifier.scopusauthoridGoyal, R=7202793341en_US
dc.identifier.scopusauthoridStrickberger, SA=7005045293en_US
dc.identifier.scopusauthoridMorady, F=35431764400en_US
dc.identifier.issnl0002-9149-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats