Article: Fast calibration method of 4 × 4 detector module for F-SPET

TitleFast calibration method of 4 × 4 detector module for F-SPET
Authors
Issue Date1991
Citation
Ieee Transactions On Nuclear Science, 1991, v. 38 n. 2 pt I, p. 770-775 How to Cite?
AbstractA novel system and method of light coupling analysis and calibration for the 4 × 4 detector module of the Columbia University spherical PET (positron emission tomography unit) is discussed, and its performance is evaluated. The new features of the system include a simulated-scintillation-signal generator which effectively replaces the scintillation signal from the photomultiplier tube and a multiple-channel analog-to-digital converter for simultaneous recording of all signals of interest. A fast and reliable calibration method for the 4 × 4 detector module and front-end circuit has been developed using this calibration system. Since the proposed calibration method is based on the light coupling analysis, it is possible to use an uncollimated source for the position calibration, thereby eliminating the time-consuming procedure usually used in calibration of the position-encoded detector assembly.
Persistent Identifierhttp://hdl.handle.net/10722/154940
ISSN
2023 Impact Factor: 1.9
2023 SCImago Journal Rankings: 0.537
ISI Accession Number ID

 

DC FieldValueLanguage
dc.contributor.authorKim, YSen_US
dc.contributor.authorLee, KDen_US
dc.contributor.authorWu, EXen_US
dc.contributor.authorHilal, SKen_US
dc.contributor.authorCho, ZHen_US
dc.date.accessioned2012-08-08T08:31:13Z-
dc.date.available2012-08-08T08:31:13Z-
dc.date.issued1991en_US
dc.identifier.citationIeee Transactions On Nuclear Science, 1991, v. 38 n. 2 pt I, p. 770-775en_US
dc.identifier.issn0018-9499en_US
dc.identifier.urihttp://hdl.handle.net/10722/154940-
dc.description.abstractA novel system and method of light coupling analysis and calibration for the 4 × 4 detector module of the Columbia University spherical PET (positron emission tomography unit) is discussed, and its performance is evaluated. The new features of the system include a simulated-scintillation-signal generator which effectively replaces the scintillation signal from the photomultiplier tube and a multiple-channel analog-to-digital converter for simultaneous recording of all signals of interest. A fast and reliable calibration method for the 4 × 4 detector module and front-end circuit has been developed using this calibration system. Since the proposed calibration method is based on the light coupling analysis, it is possible to use an uncollimated source for the position calibration, thereby eliminating the time-consuming procedure usually used in calibration of the position-encoded detector assembly.en_US
dc.languageengen_US
dc.relation.ispartofIEEE Transactions on Nuclear Scienceen_US
dc.titleFast calibration method of 4 × 4 detector module for F-SPETen_US
dc.typeArticleen_US
dc.identifier.emailWu, EX:ewu1@hkucc.hku.hken_US
dc.identifier.authorityWu, EX=rp00193en_US
dc.description.naturelink_to_subscribed_fulltexten_US
dc.identifier.doi10.1109/23.289388en_US
dc.identifier.scopuseid_2-s2.0-0026142581en_US
dc.identifier.volume38en_US
dc.identifier.issue2 pt Ien_US
dc.identifier.spage770en_US
dc.identifier.epage775en_US
dc.identifier.isiWOS:A1991FK17300121-
dc.publisher.placeUnited Statesen_US
dc.identifier.scopusauthoridKim, YS=7410195659en_US
dc.identifier.scopusauthoridLee, KD=7501516856en_US
dc.identifier.scopusauthoridWu, EX=7202128034en_US
dc.identifier.scopusauthoridHilal, SK=7004804427en_US
dc.identifier.scopusauthoridCho, ZH=7006289037en_US
dc.identifier.issnl0018-9499-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats