File Download
  Links for fulltext
     (May Require Subscription)
Supplementary

Article: Comparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patients

TitleComparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patients
Authors
Issue Date2007
PublisherWiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044
Citation
Anaesthesia, 2007, v. 62 n. 5, p. 509-515 How to Cite?
AbstractA randomised cross-over study was performed in 34 patients with no evidence of airway difficulties, following induction of general anaesthesia, to compare the efficacy of the Levitan FPS scope™ (LFPS) and the single-use bougie for tracheal intubation during simulated grade IIIa laryngoscopy. Success rates for intratracheal placement of the device, and the time required for insertion and tracheal intubation were recorded. Both devices were equally successful (31/34 for the LFPS vs 29/34 for the bougie) for insertion into the glottis. The mean insertion time for the LFPS was statistically longer than that for the bougie (4.4-12.5 s) but this difference was not clinically relevant. Intubation times were similar between the two devices. Major problems hindering successful intubation using the LFPS were the presence of a narrow epiglottic-pharyngeal wall space and copious secretions. An inability to maintain the desired shape was the principal cause of failure with the bougie. © 2007 The Authors Journal compilation © 2007 The Association of Anaesthetists of Great Britain and Ireland.
Persistent Identifierhttp://hdl.handle.net/10722/147244
ISSN
2021 Impact Factor: 12.893
2020 SCImago Journal Rankings: 1.839
ISI Accession Number ID
References
Errata

 

DC FieldValueLanguage
dc.contributor.authorGreenland, KBen_US
dc.contributor.authorLiu, Gen_US
dc.contributor.authorTan, Hen_US
dc.contributor.authorEdwards, Men_US
dc.contributor.authorIrwin, MGen_US
dc.date.accessioned2012-05-29T06:01:00Z-
dc.date.available2012-05-29T06:01:00Z-
dc.date.issued2007en_US
dc.identifier.citationAnaesthesia, 2007, v. 62 n. 5, p. 509-515en_US
dc.identifier.issn0003-2409en_US
dc.identifier.urihttp://hdl.handle.net/10722/147244-
dc.description.abstractA randomised cross-over study was performed in 34 patients with no evidence of airway difficulties, following induction of general anaesthesia, to compare the efficacy of the Levitan FPS scope™ (LFPS) and the single-use bougie for tracheal intubation during simulated grade IIIa laryngoscopy. Success rates for intratracheal placement of the device, and the time required for insertion and tracheal intubation were recorded. Both devices were equally successful (31/34 for the LFPS vs 29/34 for the bougie) for insertion into the glottis. The mean insertion time for the LFPS was statistically longer than that for the bougie (4.4-12.5 s) but this difference was not clinically relevant. Intubation times were similar between the two devices. Major problems hindering successful intubation using the LFPS were the presence of a narrow epiglottic-pharyngeal wall space and copious secretions. An inability to maintain the desired shape was the principal cause of failure with the bougie. © 2007 The Authors Journal compilation © 2007 The Association of Anaesthetists of Great Britain and Ireland.en_US
dc.languageengen_US
dc.publisherWiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044en_US
dc.relation.ispartofAnaesthesiaen_US
dc.subject.meshAdulten_US
dc.subject.meshAgeden_US
dc.subject.meshAged, 80 And Overen_US
dc.subject.meshAnesthesia, Generalen_US
dc.subject.meshCross-Over Studiesen_US
dc.subject.meshDisposable Equipmenten_US
dc.subject.meshEquipment Designen_US
dc.subject.meshFemaleen_US
dc.subject.meshFiber Optic Technology - Instrumentationen_US
dc.subject.meshHumansen_US
dc.subject.meshIntubation, Intratracheal - Instrumentationen_US
dc.subject.meshLaryngoscopesen_US
dc.subject.meshMaleen_US
dc.subject.meshMiddle Ageden_US
dc.subject.meshTime Factorsen_US
dc.subject.meshTreatment Outcomeen_US
dc.titleComparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patientsen_US
dc.typeArticleen_US
dc.identifier.emailIrwin, MG:mgirwin@hku.hken_US
dc.identifier.authorityIrwin, MG=rp00390en_US
dc.description.naturelink_to_OA_fulltexten_US
dc.identifier.doi10.1111/j.1365-2044.2007.05003.xen_US
dc.identifier.pmid17448065en_US
dc.identifier.scopuseid_2-s2.0-34249058740en_US
dc.identifier.hkuros134901-
dc.relation.referenceshttp://www.scopus.com/mlt/select.url?eid=2-s2.0-34249058740&selection=ref&src=s&origin=recordpageen_US
dc.identifier.volume62en_US
dc.identifier.issue5en_US
dc.identifier.spage509en_US
dc.identifier.epage515en_US
dc.identifier.isiWOS:000245811600015-
dc.publisher.placeUnited Kingdomen_US
dc.relation.erratumdoi:10.1111/j.1365-2044.2007.05168_62_6.x-
dc.identifier.citeulike1238692-
dc.identifier.issnl0003-2409-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats