File Download
Links for fulltext
(May Require Subscription)
- Publisher Website: 10.1111/j.1365-2044.2007.05003.x
- Scopus: eid_2-s2.0-34249058740
- PMID: 17448065
- WOS: WOS:000245811600015
- Find via
Supplementary
- Citations:
- Appears in Collections:
Article: Comparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patients
Title | Comparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patients |
---|---|
Authors | |
Issue Date | 2007 |
Publisher | Wiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044 |
Citation | Anaesthesia, 2007, v. 62 n. 5, p. 509-515 How to Cite? |
Abstract | A randomised cross-over study was performed in 34 patients with no evidence of airway difficulties, following induction of general anaesthesia, to compare the efficacy of the Levitan FPS scope™ (LFPS) and the single-use bougie for tracheal intubation during simulated grade IIIa laryngoscopy. Success rates for intratracheal placement of the device, and the time required for insertion and tracheal intubation were recorded. Both devices were equally successful (31/34 for the LFPS vs 29/34 for the bougie) for insertion into the glottis. The mean insertion time for the LFPS was statistically longer than that for the bougie (4.4-12.5 s) but this difference was not clinically relevant. Intubation times were similar between the two devices. Major problems hindering successful intubation using the LFPS were the presence of a narrow epiglottic-pharyngeal wall space and copious secretions. An inability to maintain the desired shape was the principal cause of failure with the bougie. © 2007 The Authors Journal compilation © 2007 The Association of Anaesthetists of Great Britain and Ireland. |
Persistent Identifier | http://hdl.handle.net/10722/147244 |
ISSN | 2023 Impact Factor: 7.5 2023 SCImago Journal Rankings: 2.400 |
ISI Accession Number ID | |
References | |
Errata |
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Greenland, KB | en_US |
dc.contributor.author | Liu, G | en_US |
dc.contributor.author | Tan, H | en_US |
dc.contributor.author | Edwards, M | en_US |
dc.contributor.author | Irwin, MG | en_US |
dc.date.accessioned | 2012-05-29T06:01:00Z | - |
dc.date.available | 2012-05-29T06:01:00Z | - |
dc.date.issued | 2007 | en_US |
dc.identifier.citation | Anaesthesia, 2007, v. 62 n. 5, p. 509-515 | en_US |
dc.identifier.issn | 0003-2409 | en_US |
dc.identifier.uri | http://hdl.handle.net/10722/147244 | - |
dc.description.abstract | A randomised cross-over study was performed in 34 patients with no evidence of airway difficulties, following induction of general anaesthesia, to compare the efficacy of the Levitan FPS scope™ (LFPS) and the single-use bougie for tracheal intubation during simulated grade IIIa laryngoscopy. Success rates for intratracheal placement of the device, and the time required for insertion and tracheal intubation were recorded. Both devices were equally successful (31/34 for the LFPS vs 29/34 for the bougie) for insertion into the glottis. The mean insertion time for the LFPS was statistically longer than that for the bougie (4.4-12.5 s) but this difference was not clinically relevant. Intubation times were similar between the two devices. Major problems hindering successful intubation using the LFPS were the presence of a narrow epiglottic-pharyngeal wall space and copious secretions. An inability to maintain the desired shape was the principal cause of failure with the bougie. © 2007 The Authors Journal compilation © 2007 The Association of Anaesthetists of Great Britain and Ireland. | en_US |
dc.language | eng | en_US |
dc.publisher | Wiley-Blackwell Publishing Ltd. The Journal's web site is located at http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2044 | en_US |
dc.relation.ispartof | Anaesthesia | en_US |
dc.subject.mesh | Adult | en_US |
dc.subject.mesh | Aged | en_US |
dc.subject.mesh | Aged, 80 And Over | en_US |
dc.subject.mesh | Anesthesia, General | en_US |
dc.subject.mesh | Cross-Over Studies | en_US |
dc.subject.mesh | Disposable Equipment | en_US |
dc.subject.mesh | Equipment Design | en_US |
dc.subject.mesh | Female | en_US |
dc.subject.mesh | Fiber Optic Technology - Instrumentation | en_US |
dc.subject.mesh | Humans | en_US |
dc.subject.mesh | Intubation, Intratracheal - Instrumentation | en_US |
dc.subject.mesh | Laryngoscopes | en_US |
dc.subject.mesh | Male | en_US |
dc.subject.mesh | Middle Aged | en_US |
dc.subject.mesh | Time Factors | en_US |
dc.subject.mesh | Treatment Outcome | en_US |
dc.title | Comparison of the Levitan FPS Scope™ and the single-use bougie for simulated difficult intubation in anaesthetised patients | en_US |
dc.type | Article | en_US |
dc.identifier.email | Irwin, MG:mgirwin@hku.hk | en_US |
dc.identifier.authority | Irwin, MG=rp00390 | en_US |
dc.description.nature | link_to_OA_fulltext | en_US |
dc.identifier.doi | 10.1111/j.1365-2044.2007.05003.x | en_US |
dc.identifier.pmid | 17448065 | en_US |
dc.identifier.scopus | eid_2-s2.0-34249058740 | en_US |
dc.identifier.hkuros | 134901 | - |
dc.relation.references | http://www.scopus.com/mlt/select.url?eid=2-s2.0-34249058740&selection=ref&src=s&origin=recordpage | en_US |
dc.identifier.volume | 62 | en_US |
dc.identifier.issue | 5 | en_US |
dc.identifier.spage | 509 | en_US |
dc.identifier.epage | 515 | en_US |
dc.identifier.isi | WOS:000245811600015 | - |
dc.publisher.place | United Kingdom | en_US |
dc.relation.erratum | doi:10.1111/j.1365-2044.2007.05168_62_6.x | - |
dc.identifier.citeulike | 1238692 | - |
dc.identifier.issnl | 0003-2409 | - |