Article: Inter- and intra-observer errors in identifying the transepicondylar axis and Whiteside's line.

TitleInter- and intra-observer errors in identifying the transepicondylar axis and Whiteside's line.
Authors
Issue Date2008
PublisherHong Kong Academy of Medicine Press. The Journal's web site is located at http://www.josonline.org/
Citation
Journal Of Orthopaedic Surgery (Hong Kong), 2008, v. 16 n. 3, p. 316-320 How to Cite?
AbstractPURPOSE: To assess inter- and intra-observer errors in identifying the transepicondylar axis and Whiteside's line in a cadaveric model mimicking total knee arthroplasty. METHODS: Four cadaveric knees with intact soft tissues were used. The knees were exposed anteriorly using the Insall approach, with the patella everted laterally. Three observers (2 surgeons and one trainee) took turns to identify the anatomic landmarks of the transepicondylar axis and Whiteside's line. Each observer repeated the process 20 times. Each identification was photographed and referenced with the true values obtained from the knees after they were stripped of all soft tissue. Inter- and intra-observer errors in the anatomic landmarks were compared. RESULTS: Inter-observer error was significant with both the transepicondylar axis and Whiteside's line (p<0.001, one-way ANOVA). The intra-observer variation was greater for Whiteside's line than the transepicondylar axis (standard deviation, 4.2 vs 2.5 degrees). The maximum potential errors in the transepicondylar axis and Whiteside's line were 13 degrees and 24 degrees, respectively. CONCLUSION: The accuracy of rotational alignment of the transepicondylar axis and Whiteside's line were operator-dependent, and their intra-operative reproducibility was low.
Persistent Identifierhttp://hdl.handle.net/10722/125191
ISSN
2023 Impact Factor: 1.3
2023 SCImago Journal Rankings: 0.557

 

DC FieldValueLanguage
dc.contributor.authorYan, CHen_HK
dc.contributor.authorYau, WPen_HK
dc.contributor.authorNg, TPen_HK
dc.contributor.authorLie, WHen_HK
dc.contributor.authorChiu, KYen_HK
dc.contributor.authorTang, WMen_HK
dc.date.accessioned2010-10-31T11:16:38Z-
dc.date.available2010-10-31T11:16:38Z-
dc.date.issued2008en_HK
dc.identifier.citationJournal Of Orthopaedic Surgery (Hong Kong), 2008, v. 16 n. 3, p. 316-320en_HK
dc.identifier.issn1022-5536en_HK
dc.identifier.urihttp://hdl.handle.net/10722/125191-
dc.description.abstractPURPOSE: To assess inter- and intra-observer errors in identifying the transepicondylar axis and Whiteside's line in a cadaveric model mimicking total knee arthroplasty. METHODS: Four cadaveric knees with intact soft tissues were used. The knees were exposed anteriorly using the Insall approach, with the patella everted laterally. Three observers (2 surgeons and one trainee) took turns to identify the anatomic landmarks of the transepicondylar axis and Whiteside's line. Each observer repeated the process 20 times. Each identification was photographed and referenced with the true values obtained from the knees after they were stripped of all soft tissue. Inter- and intra-observer errors in the anatomic landmarks were compared. RESULTS: Inter-observer error was significant with both the transepicondylar axis and Whiteside's line (p<0.001, one-way ANOVA). The intra-observer variation was greater for Whiteside's line than the transepicondylar axis (standard deviation, 4.2 vs 2.5 degrees). The maximum potential errors in the transepicondylar axis and Whiteside's line were 13 degrees and 24 degrees, respectively. CONCLUSION: The accuracy of rotational alignment of the transepicondylar axis and Whiteside's line were operator-dependent, and their intra-operative reproducibility was low.en_HK
dc.languageengen_HK
dc.publisherHong Kong Academy of Medicine Press. The Journal's web site is located at http://www.josonline.org/en_HK
dc.relation.ispartofJournal of orthopaedic surgery (Hong Kong)en_HK
dc.rightsJournal Of Orthopaedic Surgery. Copyright © Hong Kong Academy of Medicine Press.-
dc.subject.meshArthroplasty, Replacement, Knee-
dc.subject.meshCadaver-
dc.subject.meshFemur-
dc.subject.meshMedical Errors-
dc.subject.meshOsteoarthritis, Knee - diagnosis - surgery-
dc.titleInter- and intra-observer errors in identifying the transepicondylar axis and Whiteside's line.en_HK
dc.typeArticleen_HK
dc.identifier.openurlhttp://library.hku.hk:4550/resserv?sid=HKU:IR&issn=1022-5536&volume=16&issue=3&spage=316&epage=320&date=2008&atitle=Inter-observer+and+intra-observer+errors+in+obtaining+transepicondylar+axis+and+whiteside+line+in+total+knee+arthroplastyen_HK
dc.identifier.emailYan, CH:yanchoi@hku.hken_HK
dc.identifier.emailYau, WP:peterwpy@hkucc.hku.hken_HK
dc.identifier.emailChiu, KY:pkychiu@hkucc.hku.hken_HK
dc.identifier.authorityYan, CH=rp00303en_HK
dc.identifier.authorityYau, WP=rp00500en_HK
dc.identifier.authorityChiu, KY=rp00379en_HK
dc.description.naturelink_to_OA_fulltexten_US
dc.identifier.pmid19126898-
dc.identifier.scopuseid_2-s2.0-65449140881en_HK
dc.identifier.hkuros175956en_HK
dc.identifier.volume16en_HK
dc.identifier.issue3en_HK
dc.identifier.spage316en_HK
dc.identifier.epage320en_HK
dc.publisher.placeHong Kongen_HK
dc.identifier.scopusauthoridYan, CH=35764094100en_HK
dc.identifier.scopusauthoridYau, WP=7005822441en_HK
dc.identifier.scopusauthoridNg, TP=24438193400en_HK
dc.identifier.scopusauthoridLie, WH=26534696300en_HK
dc.identifier.scopusauthoridChiu, KY=7202988127en_HK
dc.identifier.scopusauthoridTang, WM=7403430820en_HK
dc.identifier.issnl1022-5536-

Export via OAI-PMH Interface in XML Formats


OR


Export to Other Non-XML Formats